5.1.2
Mounting location
Note the following when erecting the device:
‣
If mounting on a wall, make sure that the wall has sufficient load-bearing capacity and is
fully perpendicular.
‣
If mounting on a base, erect the device on a level surface.
‣
Protect the device against additional heating (e.g. from a heating system).
‣
Protect the device against mechanical vibrations.
‣
Protect the device against corrosive gases, e.g. hydrogen sulfide (H
2
S) .
‣
Make sure to pay attention to the maximum height difference and the maximum distance
from the sampling point.
‣
Ensure that the unit can drain freely, without any siphoning effects.
‣
Make sure air can circulate freely at the front of the housing.
‣
Open analyzers (i.e. analyzers that are supplied without a door) may only be erected in
closed areas or in a protective cabinet or similar facility.