Diagnostics and troubleshooting
Liquiline System CA80HA
156
Hauser
No.
Message
Factory settings
Tests or remedial action
S
D
F
337
Pump hose warning
M
On
Off
End of pump hose service life will be reached
shortly
Display under
Menu
/
Diagnostics
/
Operating time
information
/
Pump tube life
1.
Schedule replacement.
2.
After replacing, reset the operating time
under
Menu
/
Diagnostics
/
Operating time
information
.
360
Cooling/Heating
C
On
Off
Temperature range in the housing exceeded
1.
Check installation conditions and the
ambient temperature
2.
Replace the actuator module.
3.
Contact the Service Department.
361
Cooling/Heating
F
On
On
Cooling/heating module defective
Defined temperature range not reached. This could
affect the functionality of the reagents.
1.
Check whether the insulation cap is seated
correctly over the reagents.
2.
Replace cooling/heating module.
3.
Contact the Service Department.
362
Photometer temp.
F
On
Off
Temperature of the photometer too high
‣
Contact the Service Department.
363
Photometer temp.
F
On
Off
Temperature of the photometer too low
‣
Contact the Service Department.
364
Timeout dosing
F
On
On
The Liquid Manager/dosing dispenser has timed
out.
Possible reasons:
• Photoelectric barrier defective
• Blocking
‣
Contact the Service Department.
365
Photometer comm.
F
On
On
Photometer not communicating
Possible reasons:
Incorrect photometer connection
1.
Check photometer connection.
2.
Contact the Service Department.
367
Module connect.
F
On
On
No communication to sample preparation
‣
Check the connecting cable to the sample
preparation system.