Liquiline System CA80HA
Operation
Hauser
89
Menu/Setup/Sample preparation/Installation type = Others/Sample preparation 1 ... 2
Function
Options
Info
Cleaning duration
Selection
10 to 60 s
Factory setting
30 s
Cleaning duration of the automatic compressed air cleaning
system
CAT820 and CAT860
Heatings
Cabinet
Selection
• Available
• Not available
Factory setting
Depends on device version
You can enable the heating here with a heated sample
preparation.
Hose filter
Selection
• Available
• Not available
Factory setting
Depends on the device version
With heated sample preparation
Hose analyzer
Selection
• Available
• Not available
Factory setting
Depends on the device version
With heated sample preparation
Diagnostics settings
Limit change filter
Function
Selection
• On
• Off
Factory setting
Off
Warning limit
Selection
01
‐
00 to 99
‐
00 (DD
‐
HH)
Factory setting
60-00
Diag. code 729
Limit change hose
Function
Selection
• On
• Off
Factory setting
On