Liquiline System CA80HA
Maintenance
Hauser
177
Interval
Maintenance work
12 months
• (If necessary) replace hoses:
Norprene, black
• Replace filter mats
• Replace the O-ring in the cover of the sample collecting vessel
2 or 4 years
Measuring interval < 1 hour:
Replace Liquid Manager every 2 years
Measuring interval ≥ 1 hour:
Replace Liquid Manager every 4 years
2 to 3 years
(If necessary) replace hoses:
• C-Flex, white ID 3.2 mm
• C-Flex, white ID 6.4 mm
If required
• Clean the housing
• Check the fan of the cooling module for dirt and vacuum clean if necessary
• Clean CAT8x0 sample preparation (automatic cleaning)
• Rinse the system
12.2 Maintenance tasks
12.2.1 Software menu Maintenance
Menu/Operation/Maintenance
• Bottle replacement
If level monitoring is activated: specify when bottles are replaced.
• Dispenser replacement
Operating time counter, can be reset here. Replace dispensers: → 183
• Liquid manager replacement
Operating time counter, can be reset here. Replace Liquid Manager: → 184
• Filter mats replacement
Operating time counter, can be reset here. Replace filter mats: → 182
• Sample preparation 1 ... 2
Operating time counter, can be reset here. Operating hours of:
• Filter
• Pump hose
• Replace filter and hose: → see the Operating Instructions for the sample preparation
system
• Decommissioning
12.2.2 Setting the manual mode
Initial situation: The device is still in the automatic mode and you want to perform
maintenance work.
1. Press the
MODE
soft key.
The current mode is displayed:
Automatic