Diagnostics and troubleshooting
Liquiline System CA80HA
160
Hauser
No.
Message
Factory settings
Tests or remedial action
S
D
F
726
Liquids warning
M
On
Off
Consumable liquids, warning
Measurement can still take place.
Possible reasons
• The level of one or more liquids is low
• One or more liquids almost past their shelf life.
‣
Top up/replace the liquids in question and reset
the counter in
Diagnostics
/
Operating time
information
.
727
Liquids alarm
F
On
Off
Consumable liquids, alarm
Measurement can still take place.
Possible reasons
• The level of one or more liquids is low
• One or more liquids past their shelf life.
‣
Top up/replace the liquids in question and reset
the counter in
Diagnostics
/
Operating time
information
.
731
Leakage detected
F
On
Off
Leak detected in the sample preparation system or
analyzer
1.
Check hoses and connections.
2.
Check solenoid valves.
3.
Check leak sensor.
4.
Check the free outlet on the analyzer.
5.
Replace defective parts and, if necessary,
reset the operating hours counters in the
Diagnostics menu.
732
Wear parts alarm
F
On
On
One or more wear parts has exceeded its service
life.
‣
Replace the wear parts in question and reset the
counter in
Diagnostics
/
Operating time
information
.
733
Wear parts warning
M
On
Off
One or more wear parts has almost reached the end
of its service life.
‣
Replace the wear parts in question and reset the
counter in
Diagnostics
/
Operating time
information
.
910
Limit switch
S
On
Off
Limit switch activated
930
No sample
F
On
On
Sample flow interrupted during intake
• Suction line blocked or leaking
• No inflow of sample
1.
Check the suction line and suction strainer.
2.
Check inflow of sample.