Operation
Liquiline System CA80HA
84
Hauser
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Bottles
Function
Options
Info
Alarm limits
Diag. code 727
Cleaner C
1 to 7 d
Factory setting
2 d
Reagent RK
1 to 7 d
Factory setting
2 d
Standard S1
1 to 7 d
Factory setting
2 d
Absorption curve recording
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Absorption curve recording
Function
Options
Info
In automatic mode
Selection
• Off
• On
Factory setting
Off
Off
: Recording is only possible in the manual
mode
On
: Recording also takes place in the automatic
mode
Curve
Selection
1 to 7
Factory setting
3
Select the absorption curve to be recorded.
Only one curve can be selected at any one time.
The data are saved in a logbook.
Behavior after power failure
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
After power failure
Selection
• Last mode
• Manual mode
Factory setting
Last mode
Setting specifying how the analyzer should
behave after a power failure, or when the
power is back on.
Last mode
: The analyzer remains in the mode
that was last set. Example: The automatic
mode was set. The analyzer continues after
initialization and after any samples have been
discarded.
Manual mode
: The analyzer switches to the
manual mode and waits for user action.