Liquiline System CA80HA
Operation
Hauser
113
A message is individually assigned to the relay output at 2 specific points in the menu:
•
Menu
/
Setup
/
General settings
/
Extended setup
/
Diagnostics settings
/
Diag. behavior
(device-specific messages)
•
Menu
/
Setup
/
Inputs
/<Sensor>/
Extended setup
/
Diagnostics settings
/
Diag. behavior
(sensor-specific messages)
Before being able to assign the relay output to a special message in
Diag. behavior
you
must first configure
Outputs
/Relay x:y or /
Alarm relay
/
Function
=
Diagnostics
.
Function = Diagnostics
Function
Options
Info
Operating mode
Selection
• as assigned
• Namur M
• Namur S
• Namur C
• Namur F
Factory setting
• Relay:
as assigned
• Alarm relays:
Namur F
as assigned
If this option is selected, the diagnostic
messages which you have individually assigned
to the relay are output via the relay.
Namur M
...
Namur F
If you decided to use one of the Namur classes,
all the messages that are assigned to the
individual class are output via the relay. You
can also change the Namur class assignment
for every diagnostic message.
(
Menu
/
Setup
/
General settings
/
Extended
setup
/
Diagnostics settings
/
Diag. behavior
or
Menu
/
Setup
/
Inputs
/<Sensor>/
Extended
setup
/
Diagnostics settings
/
Diag. behavior
)
Attributed diagnostic messages
Operating mode
=
as assigned
Read only
All the messages assigned to the relay output
appear on the display. You do not have the
option of editing the information here.