Liquiline System CA80HA
Maintenance
Hauser
183
A0045364
75
Fan screen on the bottom
1. Open both fan screens and remove them.
The filter mats are located under the screens.
2. Remove the used filter mats and replace them with new ones from the maintenance kit.
3. Fit the fan screens back on.
4. Reset the operating hours counter for the mats:
Menu
/
Operation
/
Operating time
filter mats
/
Reset
.
12.2.8 Replacing dispenser
You require the following parts:
• From the CAV800 maintenance kit:
• NORPRENE hose, ID 1.6 mm
• 10 ml dispensers with adapter
• 2.5 ml dispenser with adapter
• 1 pair of gloves that are resistant to the reagents used
Preparing to replace dispensers
1. Rinse the system.→ 179
2. Remove the bottles including the reagent hoses and the bottle tray.
3. Remove the cover of the carrier board.
Removing dispensers
1. Select
Menu
/
Operation
/
Dispenser replacement
/
Dispenser selection
.
2. Select the dispensers that you want to replace.