Operation
Liquiline System CA80HA
88
Hauser
Menu/Setup/Sample preparation/Installation type = Others/Sample preparation 1 ... 2
Function
Options
Info
Operating mode
Selection
• Independent
• Controlled
Factory setting
Controlled
• Controlled
Sample preparation that is controlled by the analyzer,
e.g. Liquiline System CAT820 / CAT860
• Independent
No sample preparation that is controlled by the analyzer.
The supply of sample must be guaranteed by the
customer.
Operating mode = Independent
1.
Power supply
: Switch on if power is to be supplied
via the analyzer.
2.
Timeout no sample
: Specify the time for the
timeout (3:00 to 30:00 min:ss).
Operating mode = Controlled
‣
Make other settings as described below.
Tag
Customized text, 32 characters
Timeout no sample
3:00 to 30:00 min:ss
CAT820 and CAT860
Sampling pump interval
Pump time
Selection
10 to 20 s
Factory setting
10 s
Ratio between the pumping duration and the pumping
interval of the peristaltic pump. This affects the volume of
the pumped sample.
Pump pause
Selection
20 to 50 s
Factory setting
30 s
CAT820 and CAT860
Pressurized air cleaning
CAT820 only
Compressed air
Selection
• Available
• Not available
Factory setting
Depends on the device version
Settings for sample preparation with or without
compressed air cleaning.
In the event of retrofitting, the compressed air can be
activated here.
Cleaning mode
Selection
• On
• Off
Factory setting
On
Activates or deactivates automatic compressed air cleaning
of the pump-to-filter hose and of the filter.
Cleaning interval
Selection
0:30 to 4:00 HH:MM
Factory setting
2:00
Cleaning interval of the automatic compressed air cleaning
system