Operation
Liquiline System CA80HA
90
Hauser
Menu/Setup/Sample preparation/Installation type = Others/Sample preparation 1 ... 2
Function
Options
Info
Warning limit
Selection
01
‐
00 to 99
‐
00 (DD
‐
HH)
Factory setting
60-00
Diag. code 337
Reset settings
Resets all settings specific to sample preparation. All other
settings are retained.
Menu/Setup/Sample preparation/Installation type = Pressurized pipe sampling/Sample preparation 1 ... 2
Function
Options
Info
Cleaning valve
Selection
• Available
• Not available
Factory setting
Not available
If available, the valve can be activated here.
Cleaning valve
=
Available
Filter backflush
Selection
• On
• Off
Factory setting
On
Cleaning valve
=
Available
and
Filter backflush
=
On
Cleaning interval
Selection
00:10 to 02:00 HH:MM
Factory setting
00:30
Compressed air or water cleaning (optional) for
extended filter maintenance intervals
Cleaning valve
=
Available
Cleaning duration
Selection
10 to 30 s
Factory setting
10 s
Cleaning duration of the automatic water or
compressed air cleaning system
Cleaning valve
=
Available
Cleaning discard time
Selection
0 to 1800 s
Factory setting
180 s
Discard time once cleaning is complete. If water
is used as the flushing medium, this must be
replaced by a fresh sample before starting the
next measurement.
Timeout no sample
3:00 to 30:00 min:ss
Reset settings
Resets all settings specific to sample
preparation. All other settings are retained.