Liquiline System CA80HA
System integration
Hauser
45
Individual types of protection permitted for this product (impermeability (IP), electrical safety,
EMC interference immunity, Ex protection) can no longer be guaranteed if, for example :
• Covers are left off
• Different power units to the ones supplied are used
• Cable glands are not sufficiently tightened (must be tightened with 2 Nm (1.5 lbf ft) for the
permitted level of IP protection)
• Unsuitable cable diameters are used for the cable glands
• Modules are not fully secured
• The display is not fully secured (risk of moisture entering due to inadequate sealing)
• Loose or insufficiently tightened cables/cable ends
• Conductive cable strands are left in the device
6.7
Post-connection check
L
WARNING
Connection errors
The safety of people and of the measuring point is at risk! The manufacturer does not accept
any responsibility for errors that result from failure to comply with the instructions in this
manual.
‣
Put the device into operation only if you can answer
yes
to
all
the following questions.
Device condition and specifications
‣
Are the device and all the cables free from damage on the outside?
Electrical connection
‣
Are the mounted cables strain relieved?
‣
Are the cables routed without loops and cross-overs?
‣
Are the signal cables correctly connected as per the wiring diagram?
‣
Are all plug-in terminals securely engaged?
‣
Are all the connection wires securely positioned in the cable terminals?
7
System integration
7.1
Overview of device description files
7.1.1
Device identification
→ 201, protocol-specific data
7.1.2
Source for device description and device master files
Downloading device drivers
1. Go to the website:
A dropdown list is displayed.
2. Select: device driver.