Operation
Liquiline System CA80HA
86
Hauser
10.3.3 Calibration
Menu/Setup/Analyzer/Calibration
Function
Options
Info
Start condition
Selection
• Immediate
• Date/time
Factory setting
Immediate
The calibration can start either immediately or
at the set time.
If the start condition selected is
Immediate
Date
01.01.1970 to 07.02.2106
Functions only available for:
Start condition
=
Date/time
Time
00:00:00 to 23:59:59
Calibration interval
0-01 to 90-00 (
DD-hh
)
Factory setting
03-00
DD-hh
For configuring the time interval of the
calibration/adjustment.
Next calibration
Mode
=
Automatic
Read only
Zero point
Read only
Calibration factor
Read only
Relationship of the measured concentration to
the pre-specified concentration of the
calibration standard.
Settings
Nominal concentration
1.00 to 800.00 (mg/l)
Factory setting
65.00 (based on CaCO
3
)
For configuring the concentration of the
calibration standard solution.
Depends on the settings under
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Measuring value
/
Main value
Automatic cleaning
Selection
• Off
• On
Factory setting
On
For specifying whether cleaning will take place
before every calibration/adjustment (only in
the automatic mode).