Operation
Liquiline System CA80HA
70
Hauser
Menu/Setup/General settings/Logbooks
Function
Options
Info
Line plotter
Menu to define the graphic display
Axes
Selection
• Off
• On
Factory setting
On
Should the axes (x, y) be displayed (
On
) or not
(
Off
)?
Orientation
Selection
• Horizontal
• Vertical
Factory setting
Horizontal
You can choose whether the value curves
should be displayed from left to right
(
Horizontal
) or from top to bottom (
Vertical
).
If you want to display two data logbooks
simultaneously, make sure that both logbooks
have the same settings here.
X-Description
Selection
• Off
• On
Factory setting
On
‣
Decide whether a description should be
displayed for the axes and whether
gridlines should be shown. In addition, you
can also decide whether pitches should be
displayed.
Y-Description
Grids
Pitches
X Pitch/Grid distance
10 to 50%
Factory setting
10 %
‣
Determine the pitch.
Y Pitch/Grid distance
Remove
Action
This action removes the data logbook. Any
data that have not been saved are lost.
Example: New data logbook (Setup/General settings/Logbooks/Data logbooks/New)
1. Make the settings:
• Logbook name
Assign a name. Example: "01".
• Source of data
Select a data source. Example: Signal of binary input 1.
• Measured value
Select the measured value to be recorded. Example: pH value.
• Scan time
Specify the time interval between two logbook entries.
• Data logbook
Activate the logbook: specify the data storage method.
2. ../
Finished
: Perform the action.
The device shows the new logbook in the list of data logbooks.
3. Select data logbook "01".
Additional display:
Log time left
.