background image

Liquiline System CA80HA

Operation

Hauser

93

10.6.1   Application examples

Chlorine regulation with feedforward control

CCA250

CA80

Flow, e.g. Promag53

A

BU

BK

BN

CCA250

INS

47

48

91

92

CA80

DIO

1

2

3

Dig.in

CA80: DIO

Dig.in: PFM

91

92

+

45

46

CA80

DIO

Dig.out: PFM

  A0029239

 60

Example of chlorine control with feedforward control

1

Connection of the inductive proximity switch INS of CCA250 to the digital input of the DIO module

2

Connection of the signal from a flowmeter to the digital input of the DIO module

3

Activation of a (pulse) dosing pump via the digital output of the DIO module

A

Dosing pump

Utilize the advantage of the effectively wear-free control with binary outputs versus a control

system with relay. With pulse frequency modulation (PFM), it is possible to achieve virtually

continuous dosing using a dosing pump with higher input frequency.

1. Connect the INS proximity switch of the CCA250 assembly to a digital input of the DIO

module (e.g. slot 6, port 1).

Summary of Contents for Liquiline System CA80HA

Page 1: ...Products Solutions Services Operating Instructions Liquiline System CA80HA Colorimetric analyzer for total hardness BA01772C 07 EN 04 21 71532810 2021 01 31 Valid as of version 01 09 00 ...

Page 2: ......

Page 3: ...tegrating the measuring device into the system 46 8 Operation options 51 8 1 Structure and function of the operating menu 51 8 2 Access to the operating menu via the local display 51 9 Commissioning 54 9 1 Preparatory steps 55 9 2 Function check 58 9 3 Switching on the measuring device 59 9 4 Setting the operating language 59 9 5 Configuring the measuring device 59 10 Operation 60 10 1 Display 60 ...

Page 4: ...Device specific accessories 189 14 2 Communication specific accessories 195 14 3 System components 196 15 Technical data 197 15 1 Input 197 15 2 Output 198 15 3 Current outputs active 199 15 4 Relay outputs 200 15 5 Protocol specific data 201 15 6 Power supply 204 15 7 Performance characteristics 206 15 8 Environment 207 15 9 Process 208 15 10 Mechanical construction 208 Index 210 ...

Page 5: ... Consequences of non compliance if applicable Corrective action This symbol alerts you to a dangerous situation Failure to avoid this situation can result in minor or more serious injuries NOTICE Cause situation If necessary Consequences of non compliance if applicable Action note This symbol alerts you to situations which may result in damage to property 1 2 Symbols Additional information tips Pe...

Page 6: ...rief Operating Instructions Liquiline System CA80HA KA01331C Operating Instructions Memosens BA01245C Software description for Memosens inputs Calibration of Memosens sensors Sensor specific diagnostics and troubleshooting Guidelines for communication via fieldbus and web server PROFIBUS SD01188C Modbus SD01189C Web server SD01190C EtherNet IP SD01293C Special documentation on reagents CY80HA SD02...

Page 7: ...uid media The analyzer is designed for use in the following applications Optimization of reverse osmosis systems and ion exchangers Classification of the level of hardness of drinking water Guaranteeing process water quality in production plants Use of the device for any purpose other than that described poses a threat to the safety of people and of the entire measuring system and is therefore not...

Page 8: ... rinse with clear water Then dry the cleaned areas with a cloth LCAUTION Risk of injury from door stop mechanism Always open the door fully to ensure the door stop engages properly 2 5 Product safety 2 5 1 State of the art technology The product is designed to meet state of the art safety requirements has been tested and left the factory in a condition in which it is safe to operate The relevant r...

Page 9: ...nal 2 Window 3 Air vent for cooling 4 Bottle tray for reagent and standard 5 Reagent insulation box optional 6 Photometer 7 Controller 3 2 Measuring system A complete measuring system comprises Analyzer Liquiline System CA80HA with the configuration ordered Integrated cooling module to extend the reagent life time and standard solution to be ordered separately Sample conditioning Liquiline System ...

Page 10: ...µm or 0 4 µm Communication via Memosens protocol control via CA80 Backflushing with compressed air version with Memosens technology Easy installation with Flexdip CYH112 TI00430C Application aeration surface water Membrane filtration Liquiline System CAT860 Function sampling and filtration Ceramic membrane filter candle pore size 0 1 µm or 0 4 µm Communication via Memosens protocol control via CA8...

Page 11: ...ting vessel overflow 4 Sample 5 Pressurized sample 6 Filter unit 7 Liquiline System CAT810 1 2 3 4 5 6 7 8 9 A0044813 4 Measuring system with Liquiline System CAT810 and cleaning valve 1 Overflow 2 Liquiline System CA80 3 Sample collecting vessel overflow 4 Sample 5 Cleaning valve 6 Pressurized sample 7 Filter unit 8 Purge connection compressed air or water 9 Liquiline System CAT810 ...

Page 12: ...System CA80HA 12 Endress Hauser 1 2 4 5 3 6 7 A0044810 5 Measuring system with Liquiline System CAT820 1 Pump 2 Liquiline System CA80 3 Sample collecting vessel 4 Sample collecting vessel overflow 5 Sample 6 Filter ceramic 7 Medium ...

Page 13: ...uiline System CA80HA Product description Endress Hauser 13 1 2 3 4 5 6 A0044811 6 Measuring system with 2x Liquiline System CAT820 1 Pump 2 Liquiline System CA80 3 Sample 4 Valve 5 Medium 6 Filter ceramic ...

Page 14: ... 8 9 A0044808 7 Measuring system with Liquiline System CA80 Liquiline System CAT820 and second analyzer 1 Backflushing with compressed air optional 6 Second analyzer 2 Valve optional 7 Sample to second analyzer 3 Pump 8 Filter ceramic 4 Liquiline System CA80 9 Medium 5 Sample ...

Page 15: ...basic module Port 1 input 1 pH glass sensor Outputs and relays are named according to their function e g current output and are displayed in ascending order with the slot and port numbers Display shows SP1 analyzer measuring channel 1 with sampling point SP1 measured value display is parameter specific is not illustrated in the example 3 3 2 Terminal diagram The unique terminal name is derived fro...

Page 16: ...he issue has been resolved 2 Verify that the contents are undamaged Notify the supplier of any damage to the delivery contents Keep the damaged goods until the issue has been resolved 3 Check that the delivery is complete and nothing is missing Compare the shipping documents with your order 4 Pack the product for storage and transportation in such a way that it is protected against impact and mois...

Page 17: ...fety information and warnings Certificate information Approvals as per order version Compare the information on the nameplate with the order 4 2 2 Product identification Product page www endress com ca80ha Interpreting the order code The order code and serial number of your product can be found in the following locations On the nameplate In the delivery papers Obtaining information on the product ...

Page 18: ... EU directives The manufacturer confirms successful testing of the product by affixing to it the mark 4 4 2 Other standards and guidelines cCSAus The product meets the requirements as per CLASS 2252 06 Process Control Equipment and CLASS 2252 86 Process Control Equipment It is tested to Canada and USA standards CAN CSA C22 2 No 61010 1 12 UL Std No 61010 1 3rd Edition EAC The product has been cert...

Page 19: ...ice can be installed in the following ways Mounted on a wall Mounted on a base 5 1 1 Dimensions 530 20 87 463 18 23 793 31 22 A0028820 12 Liquiline System CA80 closed version dimensions in mm in 417 16 42 793 31 22 530 20 87 A0030419 13 Liquiline System CA80 open version dimensions in mm in ...

Page 20: ...Installation Liquiline System CA80HA 20 Endress Hauser 930 36 61 654 25 74 1723 67 83 A0028821 14 Liquiline System CA80 with base dimensions in mm in ...

Page 21: ...ce Protect the device against additional heating e g from a heating system Protect the device against mechanical vibrations Protect the device against corrosive gases e g hydrogen sulfide H2S Make sure to pay attention to the maximum height difference and the maximum distance from the sampling point Ensure that the unit can drain freely without any siphoning effects Make sure air can circulate fre...

Page 22: ...uired for installing analyzer 330 12 99 100 3 94 A0036774 16 Minimum spacing required for mounting Engineering unit mm in 116 A0036775 17 Maximum opening angle Spacing required for installing wall mount version 750 29 53 120 4 72 Ø 11 0 43 Ø 10 7 0 42 A0036779 18 Holder unit dimensions Engineering unit mm in ...

Page 23: ...k that the analyzer is fully hooked into the wall holder unit at the top and bottom and secure the analyzer to the upper wall holder unit using the securing screw The mounting materials required to secure the device to the wall are not supplied 1 Provide the mounting materials to secure the device to the wall screws wall plugs onsite 2 Mount the wall holder unit 2 parts on the wall 3 Secure the mo...

Page 24: ...l holder unit 1 5 Fix the mount and wall holder unit in place with the screw supplied 2 5 2 2 Installing version with analyzer stand LCAUTION Incorrect installation can cause injury and damage the device If using the version with analyzer stand make sure that the analyzer stand is secured to the floor ...

Page 25: ...s 4 x M10 Dimensions of Liquiline System CA80 A0036785 20 Securing the base 1 Screw the base to the ground 2 With 2 people lift the analyzer and fit it on the base Use the recessed grips 3 Secure the base to the analyzer using the 6 screws supplied 5 3 Post installation check After mounting check all the connections to ensure they are secure ...

Page 26: ...le length 4 3 m 14 1 ft Order version CA80xx CA CSA C US General Purpose Power supply cable as per North American standard Mains voltage The maximum mains voltage fluctuation may not be more than 10 of the values indicated on the nameplate Analog signal and transmission lines e g LiYY 10 x 0 34 mm2 6 2 Connecting the analyzer NOTICE The device does not have a power switch You must install the devi...

Page 27: ...Memosens sensors or digital fieldbuses 1 Remove the bottle tray Lift up the recessed grip slightly and pull it towards the front 2 Remove the cover that is hooked into place 3 Remove all liquid bearing suction lines from the Liquid Manager 4 6 x Release the 6 screws of the carrier board using an Allen key ...

Page 28: ...ards the front 7 Only for order versions with G or NPT glands Replace the pre installed M thread cable glands with the G or NPT cable glands that are enclosed This does not affect the M32 hose glands 8 Guide the cables through the cable glands on the bottom of the device For all versions 9 Route the cables on the rear panel of the device so that they are properly protected Use cable clips 10 Guide...

Page 29: ... case of devices with a 24 V power supply the connection cross section must be at least 2 5 mm2 and may not exceed 4 mm2 With 24V power supply a current of up to 10A can flow For this reason pay attention to the voltage drop on the supply line The voltage at the device terminals must be within the specified range 204 1 To access the electronics compartment proceed as described in the Routing the c...

Page 30: ...onal ground The labeling on the terminal block is selected in such a way that it applies both for 24 V versions and and for the other device versions L and N 6 3 Connecting sample preparation 6 3 1 Connecting the optional cleaning valve of Liquiline System CAT810 1 Disconnect the mains plug 2 To fold out the carrier board towards the front proceed as described in the Routing the cables section ...

Page 31: ... A0044866 7 Connect the cleaning valve to the following plug in terminals SP1 N L PE 86 85 97 98 BK BK GN YE GY PK GN YE 1 2 A0028926 22 Connection scheme for Liquiline System CAT810 1 Liquiline System CAT810 100 to 120 V 200 to 240 V AC 2 Are not used 8 Secure the protective cover after connecting Make sure that no cables or hoses are jammed 9 Use the 6 screws to secure the carrier board after co...

Page 32: ...ening the screw 2 Nm 1 Disconnect the mains plug 2 To fold out the carrier board towards the front proceed as described in the Routing the cables section 3 Release a suitable hose gland on the right hand underside of the analyzer and remove the dummy plug from the gland 4 Guide the spiral hose through the hose gland 5 Order versions with G and NPT cable glands replace the pre mounted M thread cabl...

Page 33: ...1 2 1 2 A0028924 26 Connection scheme for Liquiline System CAT820 860 1 Hose heater 100 to 120 V 200 to 240 V AC optional 2 Connections for Memosens and communication with analyzer optional SP1 Sample inlet SP2 2nd sample inlet optional 11 Secure the protective cover after connecting Make sure that no cables or hoses are jammed 12 Use the 6 screws to secure the carrier board after connecting 6 4 C...

Page 34: ...connection Liquiline System CA80HA 34 Endress Hauser 1 2 3 4 BASE2 E 1 2 1 2 1 2 A0044867 27 Connection compartment in the controller housing 1 Base module Base2 E 2 Analyzer interface 3 Dummy cover 4 Module cover ...

Page 35: ...for display cable 1 4 Ethernet interface 5 Service interface 6 Connections for 2 Memosens sensors 7 Current outputs 8 Socket for internal power supply cable 1 9 Alarm relay connection 1 Internal device connection Do not disconnect the plug 6 4 2 Connecting the sensors Only use terminated original cables where possible G N Y E YE GN BN WH GND Com A Com B GY A0024019 29 Example of a Memosens CYK10 d...

Page 36: ... shock Change or extend the hardware for the non hazardous area always fill the slots from left to right Do not leave any gaps If all of the slots are not occupied in the case of devices for the non hazardous area always insert a dummy cover or end cover in the slot to the right of the last module 27 34 This ensures that the unit is shock protected Always ensure shock protection is guaranteed part...

Page 37: ...Liquiline System CA80HA Electrical connection Endress Hauser 37 Digital inputs and outputs DIO module 31 Module 47 47 48 48 1 2 1 2 1 2 45 46 45 46 91 92 91 92 32 Wiring diagram ...

Page 38: ...Electrical connection Liquiline System CA80HA 38 Endress Hauser Current inputs Module 2AI 23 24 23 24 33 Module 23 23 24 24 0 4 20 mA 1 2 34 Wiring diagram ...

Page 39: ...System CA80HA Electrical connection Endress Hauser 39 Current outputs 2AO 4AO 35 Module 31 31 32 32 0 4 20 mA 1 2 36 Wiring diagram 37 Module 31 31 32 32 0 4 20 mA 1 2 31 31 32 32 0 4 20 mA 3 4 38 Wiring diagram ...

Page 40: ...iquiline System CA80HA 40 Endress Hauser Relays Module 2R Module 4R 39 Module 41 43 42 Relay 1 41 43 42 Relay 2 40 Wiring diagram 41 Module 41 43 42 Relay 3 41 43 42 Relay 4 41 43 42 Relay 1 41 43 42 Relay 2 42 Wiring diagram ...

Page 41: ...only to terminals of an additional module AOR 2R or 4R not to the alarm relay of the base module Example Connecting the Chemoclean CYR10 injector cleaning unit 41 43 42 Relay 1 41 43 42 Relay 2 1 2 1 2 N L PE CM44x CM44xR CM44P CYR10 1 2 3 4 5 A0028598 43 Connecting the CYR10 injector cleaning unit 1 External power supply 2 Cleaner to spray head 3 Container with cleaner 4 Motive water 2 to 12 bar ...

Page 42: ... PROFIBUS DP Modbus RS485 95 A B 96 B A 99 Not connected C 82 DGND DGND 81 VP VP LEDs on front of module LED Identifier Color Description RJ45 LNK ACT GN RJ45 is disabled Ethernet communication is via the BASE2 module RJ45 10 100 YE PWR Power GN Supply voltage is applied and module is initialized BF Bus failure RD Bus failure SF System failure RD Device error COM Communication YE Modbus message se...

Page 43: ...eration Service OFF The switch has no function 6 5 Hardware settings 6 5 1 Bus termination module 485 only There are 2 ways to terminate the bus 1 Internal termination via DIP switch on module board ON OFF 46 DIP switch for internal termination Using a suitable tool such as a tweezer move all four DIP switches to the ON position The internal termination is used 390 W 220 W 390 W VP DGND B B A A 47...

Page 44: ...dbus If you configure an invalid address software addressing is automatically enabled via the local configuration or via the fieldbus 1 2 4 8 16 32 64 128 on off A0026776 48 Valid PROFIBUS address 67 1 2 4 8 16 32 64 128 on off A0026777 49 Valid Modbus address 195 1 2 4 8 16 32 64 128 on off A0026778 50 Invalid address 255 1 1 Order configuration software addressing is active software address conf...

Page 45: ... of people and of the measuring point is at risk The manufacturer does not accept any responsibility for errors that result from failure to comply with the instructions in this manual Put the device into operation only if you can answer yes to all the following questions Device condition and specifications Are the device and all the cables free from damage on the outside Electrical connection Are ...

Page 46: ...ter to the Ethernet port of the BASE2 module BASE2 Web Browser A0044872 51 Web server Ethernet connection Establishing the data connection All versions with the exception of PROFINET You must switch off the DHCP parameter in the Ethernet settings for your device to have a valid IP address Menu Setup General settings Extended setup Ethernet Settings You can assign the IP address manually in the sam...

Page 47: ...the Internet browser 10 If you use a proxy server to connect to the Internet Disable the proxy browser settings under Connections LAN settings 11 Enter the IP address of your device in the address bar 192 168 1 212 in the example The system takes a few moments to establish the connection and then the CM44 web server starts You might be asked for a password The factory setting is admin for the user...

Page 48: ...M required for this is an integral part of the Endress Hauser Interface Device DTM Library Download https portal endress com webdownload FieldCareDownloadGUI 7 2 2 Service interface You can connect the device to a computer via the service interface and configure it using FieldCare Furthermore configurations can also be saved transferred and documented Connection 1 Connect the service connector to ...

Page 49: ...he menu structure corresponds to the onsite operation The functions of the Liquiline soft keys are found in the main window on the left Clicking a menu name or a function corresponds to pressing the navigator You can make your settings conveniently via the computer keyboard You can use FieldCare to save logbooks make backups of configurations and transfer configurations to other devices You can al...

Page 50: ...scribed For detailed information on Modbus communication see the product pages on the Internet SD01189C EtherNet IP With the BASE2 module and the appropriate device version you can communicate via EtherNet IP Connect the EtherNet IP data cable to the RJ45 socket of the BASE2 module For detailed information on Ethernet IP communication see the product pages on the Internet SD01293C PROFINET With th...

Page 51: ...rating keys 1 Press the navigator for longer than 2 s A context menu for locking the operating keys is displayed You have the choice of locking the keys with or without password protection With password means that you can only unlock the keys again by entering the correct password Set this password here Menu Setup General settings Extended setup Data management Change key lock password 2 Select wh...

Page 52: ...y only You can only read the values but cannot change them Typical read only values are analyzer data sensor data and system information Example Menu Setup Analyzer Measuring parameter Picklists You receive a list of options In a few cases these also appear in the form of multiple choice boxes Usually you just select one option in rare instances you select one or more options Example Menu Setup Ge...

Page 53: ... loading configurations Triggering cleaning programs Example Menu Operation Manual operation User defined text You are assigning an individual designation Enter a text You can use the characters in the editor for this purpose upper case and lower case letters numbers and special characters Using the soft keys you can Cancel your entries without saving the data Delete the character in front of the ...

Page 54: ...ore the supply voltage is applied On account of the device design high switch on currents occur when the device is commissioned at low temperatures The power value indicated on the nameplate refers to the power consumption after one minute of operation when the device is commissioned at 5 C 41 F Automatic cleaning function for the suction hose Risk of injury from contact with very acidic cleaning ...

Page 55: ...glands It should not be possible to remove the hoses without the application of force 3 Visually inspect all the hose connections to ensure everything is correct Use the hose connection diagram 55 4 Insert the bottles and make the most important menu settings 59 9 1 2 Hose connection diagram A0041298 56 Hose connection diagram The diagrams below reflect the status at the time of issue of this docu...

Page 56: ...W1 Cu M SC D D5 D6 D7 D1 D2 7 C C 2 1 W2 optional dilution A0034793 57 Liquiline System CA80HA single channel device S1 Standard 1 P Sample RK Reagent RK 2 3 7 Dispensers W1 Outlet D Outlet Cu Cuvette SC Sample collecting vessel M Photometer measuring cell C Cleaner 1 T piece ...

Page 57: ...W1 Cu M D D5 D6 D7 7 C C SC SC D1 D2 2 1 W2 optional dilution A0034794 58 Liquiline System CA80HA two channel device S1 Standard 1 P Sample RK Reagent RK 2 3 7 Dispensers W1 Outlet D Outlet Cu Cuvette SC Sample collecting vessel M Photometer measuring cell C Cleaner 1 T piece ...

Page 58: ... priming system connect the intake hose supplied 1 5 m 4 92 ft to the Liquid Manager sample hose connection diagram and guide it through the hose gland of the analyzer to the outside 4 If present connect the communication cable and hose heater of the sample preparation system to the analyzer 5 Ensure that only sample that has a low solids content is supplied as otherwise there is a risk of blockag...

Page 59: ...ers inserted correctly Can the dispensers move up and down freely Are all the hose connections leak tight If sample preparation is provided has the connection been made Are the protective hoses in the hose glands strain relieved If sample preparation is not connected is the sample hose in the hose gland strain relieved Have the bottles with reagents and standard been inserted and connected 9 3 Swi...

Page 60: ...ionally connected sensors use the factory settings of the specific sensor type and the individual calibration settings that were last saved If you want to already configure additional input and output parameters in the Basic setup analyzer Configure the current outputs relays limit switches and device diagnostics with the following submenus 10 Operation 10 1 Display 10 1 1 Soft keys in the measuri...

Page 61: ... modes 2 5 Turn the navigator The display changes from channel to channel 10 1 3 Device status Icons on the display alert you to special device states Icon Location Description Header bar Diagnostic message Failure Header bar Diagnostic message Maintenance request Header bar Diagnostic message Check Header bar Diagnostic message Out of specification Header bar Fieldbus or TCP IP communication acti...

Page 62: ...ions Info Temperature unit Selection C F K Factory setting C Current output range Selection 0 20 mA 4 20 mA Factory setting 4 20 mA In accordance with Namur NE43 the linear range is from 3 8 to 20 5 mA 4 20 mA or from 0 to 20 5 mA 0 20 mA If the range is exceeded or undershot the current value stops at the range limit and a diagnostic message 460 or 461 is output Failure current 0 0 to 23 0 mA Fac...

Page 63: ...s specific fluctuations Device hold Selection Disabled Enabled Factory setting Disabled You can enable an immediate general hold for sensors here The function acts in the same way as the HOLD soft key in the screens 10 2 2 Date and time Menu Setup General settings Date Time Function Options Info Set date Depends on the format Editing mode Day two digit 01 to 31 Month two digit 01 to 12 Year four d...

Page 64: ...tory setting None None Greenwich Mean Time London DST Selection Off Europe USA Manual Factory setting Off The controller adapts the summertime normal time changeover automatically if you choose European or American daylight saving time Manual means that you can specify the start and end of daylight saving time yourself Here two additional submenus are displayed in which you specify the changeover ...

Page 65: ...vents Calibration adjustment events Operator events Diagnostic events Analyzer events Measured values You define how the logbooks should store the data In addition you can also define individual data logbooks for sensors 1 Assign the logbook name 2 Select the measured value to be recorded 3 Set the scan time Scan time You can set the scan time individually for every data logbook Further informatio...

Page 66: ...ng buffer All diagnostic messages are recorded Ring buffer If the memory is full the most recent entry automatically overwrites the oldest entry Fill up buffer If the memory is 80 full the device displays a diagnostic message If the memory is full there is an overflow i e no new values can be saved The controller displays a corresponding diagnostic message The memory then has to be cleared manuall...

Page 67: ...n Factory setting On Should the axes x y be displayed On or not Off Orientation Selection Horizontal Vertical Factory setting Horizontal You can choose whether the value curves should be displayed from left to right Horizontal or from top to bottom Vertical If you want to display two data logbooks simultaneously make sure that both logbooks have the same settings here X Description Selection Off O...

Page 68: ...values depending on the data source Scan time 0 00 01 to 1 00 00 Factory setting 0 01 00 Minimum time interval between two entries Format H MM SS Data logbook Selection Ring buffer Fill up buffer Factory setting Ring buffer Ring buffer If the memory is full the most recent entry automatically overwrites the oldest entry Fill up buffer If the memory is full there is an overflow i e no new values ca...

Page 69: ...nutes remaining until the logbook is full Log size Data logbook Fill up buffer Read only Displays the number of entries remaining until the logbook is full Logbook name Customized text 20 characters You can change the name here again Scan time 0 00 01 to 1 00 00 Factory setting 0 01 00 As above Minimum time interval between two entries Format H MM SS Data logbook Selection Ring buffer Fill up buff...

Page 70: ...hether gridlines should be shown In addition you can also decide whether pitches should be displayed Y Description Grids Pitches X Pitch Grid distance 10 to 50 Factory setting 10 Determine the pitch Y Pitch Grid distance Remove Action This action removes the data logbook Any data that have not been saved are lost Example New data logbook Setup General settings Logbooks Data logbooks New 1 Make the...

Page 71: ... only Diagnostic message Selection On Off Factory setting Depends on the message You can deactivate or reactivate a diagnostic message here Deactivating means No error message in the measuring mode No failure current at the current output Failure current Selection On Off Factory setting Depends on the message Decide whether an failure current should be output at the current output if the diagnosti...

Page 72: ... version Cleaning program for sensors Selection None Cleaning 1 Cleaning 2 Cleaning 3 Cleaning 4 Factory setting None Decide whether the diagnostic message should trigger a cleaning program You can define the cleaning programs under Menu Setup Additional functions Cleaning Detail information Read only Here you can find more information on the diagnostic message and instructions on how to resolve t...

Page 73: ...ing On You can switch off communication at this point The software can then only be accessed via local operation Termination Read only If the device is the last in the bus you can terminate via the hardware Settings Transmission Mode Selection TCP RTU ASCII Factory setting Modbus RS485 only RTU The transmission mode is displayed depending on the version ordered In the case of RS485 transmission yo...

Page 74: ...tration on and off at this point This makes it possible to create multiple users with password access User administration List of users already created View edit You can change user names or passwords or delete users A user is already created at the factory admin with password admin New user Name Free text Create new user 1 INSERT 2 Assign any name to the new user 3 Select a password for the user ...

Page 75: ...without the need for manual configuration Normally only the automatic assignment of the IP address must be configured at the client During startup the IP address the netmask and the gateway are retrieved from a DHCP server Do you want to assign the IP address of the device manually If so you have to set DHCP Off IP Address xxx xxx xxx xxx An IP address is an address in computer networks which are ...

Page 76: ...gs are used Data management Firmware update Please contact your local sales office for information on firmware updates available for your controller and its compatibility with earlier versions Current firmware version of the analyzer actuator control module Menu Diagnostics System information Back up your current setup and your logbooks to an SD card To install a firmware update you must have the ...

Page 77: ... the setup When you load a setup the current configuration is overwritten Note that cleaning and controller programs may be active Do you want to continue anyway 1 Insert the SD card into the controller card reader A setup must have been saved to the SD card 2 Go to Menu Setup General settings Extended setup Data management Load setup A list of all the setups on the SD card is displayed An error m...

Page 78: ... to other devices Activation code You require activation codes for Additional functions e g fieldbus communication Firmware upgrades If your original device has activation codes these can be found on the nameplate The corresponding device functions are activated at the factory You only require the codes if servicing the device 1 Enter the activation code Menu Setup General settings Extended setup ...

Page 79: ...ord factory setting 0000 Enter new password 2 Enter a new password Confirm new password 3 Enter the new password again The password has been changed successfully Press the navigator for longer to return to the measuring mode 10 3 Analyzer Menu Setup Analyzer Function Options Info Mode Read only Manual automatic or fieldbus Device tag Customized text 32 characters Factory setting Analyzer_serial nu...

Page 80: ...value Standard practice is to keep the measured value of the last successful measurement until the next successful measurement To detect an error during measurement the NAMUR status signal the time stamp of the measurement and the measured value status must be evaluated Status bad If a measurement cannot be performed due to an error the measured value status is set to Bad and the measured value to...

Page 81: ...ys accessible The input signal level at the binary inputs does not affect actions that require sample measurement calibration cleaning Binary input x y When the signal is active at the selected input the analyzer starts actions where sample is required Otherwise the analyzer defers or skips the action In the case of two channel devices SP2 is also displayed 1 X is context dependent is automaticall...

Page 82: ... for the remaining time until the bottle refills are consumed Menu Setup Analyzer Extended setup Diagnostics settings Bottles Function Options Info Control Selection Off On Factory setting Off On After the monitoring function is switched on you can find 2 additional menu items in Menu Operation Maintenance Change bottle Remaining operating time Bottle filling levels Bottle filling levels Start vol...

Page 83: ...ing 5 Alarm limits Diag code 727 Alarm limits are intended to immediately trigger maintenance tasks Replace the materials concerned as quickly as possible Cleaner C 1 to 20 Factory setting 2 Reagent RK 1 to 40 Factory setting 5 Standard S1 1 to 20 Factory setting 2 Remaining operating time Warning limits Diag code 726 Cleaner C 1 to 14 d Factory setting 14 d Reagent RK 1 to 14 d Factory setting 14...

Page 84: ...ding also takes place in the automatic mode Curve Selection 1 to 7 Factory setting 3 Select the absorption curve to be recorded Only one curve can be selected at any one time The data are saved in a logbook Behavior after power failure Menu Setup Analyzer Extended setup Function Options Info After power failure Selection Last mode Manual mode Factory setting Last mode Setting specifying how the an...

Page 85: ...Factory setting 0 10 Function only available for Start condition Immediate or Date time For configuring the time interval of the measurement Sequence of measurements Selection SP1 SP2 Pause Defines the sequence of channels in automatic mode Is displayed only in the case of a two channel device SP1 Indicate the number of consecutive measurements for channel SP1 SP2 Indicate the number of consecutiv...

Page 86: ...D hh For configuring the time interval of the calibration adjustment Next calibration Mode Automatic Read only Zero point Read only Calibration factor Read only Relationship of the measured concentration to the pre specified concentration of the calibration standard Settings Nominal concentration 1 00 to 800 00 mg l Factory setting 65 00 based on CaCO3 For configuring the concentration of the cali...

Page 87: ...00 DD hh Factory setting 02 00 DD hh For configuring the time interval for additional cleaning Next cleaning Mode Automatic Read only 10 4 Sample preparation The menu displayed depends on the connected sample preparation system The menu item does not appear if a Liquiline System CA80 analyzer without a collecting vessel is connected The Liquiline System CAT860 can only be operated with a Liquiline...

Page 88: ... 32 characters Timeout no sample 3 00 to 30 00 min ss CAT820 and CAT860 Sampling pump interval Pump time Selection 10 to 20 s Factory setting 10 s Ratio between the pumping duration and the pumping interval of the peristaltic pump This affects the volume of the pumped sample Pump pause Selection 20 to 50 s Factory setting 30 s CAT820 and CAT860 Pressurized air cleaning CAT820 only Compressed air S...

Page 89: ...pends on device version You can enable the heating here with a heated sample preparation Hose filter Selection Available Not available Factory setting Depends on the device version With heated sample preparation Hose analyzer Selection Available Not available Factory setting Depends on the device version With heated sample preparation Diagnostics settings Limit change filter Function Selection On ...

Page 90: ...Filter backflush Selection On Off Factory setting On Cleaning valve Available and Filter backflush On Cleaning interval Selection 00 10 to 02 00 HH MM Factory setting 00 30 Compressed air or water cleaning optional for extended filter maintenance intervals Cleaning valve Available Cleaning duration Selection 10 to 30 s Factory setting 10 s Cleaning duration of the automatic water or compressed air...

Page 91: ...re starting the next measurement Lead time 0 00 to 30 00 min ss Factory setting 25 00 min Duration 0 00 to 60 00 min ss Factory setting 25 00 min Timeout no sample 3 00 to 30 00 min ss Reset settings Resets all settings specific to sample preparation All other settings are retained 10 5 Current inputs The input can be used as the data source for limit switches and logbooks for example Furthermore ...

Page 92: ...t Factory setting 10 0 engineering unit Damping 0 to 60 s Factory setting 0 s The damping causes a floating average curve of the measured values over the time specified 1 x y slot no input number 10 6 Binary inputs and outputs Hardware options such as the DIO module with 2 digital inputs and 2 digital outputs or fieldbus module 485 enable the following Via a digital input signal measuring range sw...

Page 93: ...CCA250 to the digital input of the DIO module 2 Connection of the signal from a flowmeter to the digital input of the DIO module 3 Activation of a pulse dosing pump via the digital output of the DIO module A Dosing pump Utilize the advantage of the effectively wear free control with binary outputs versus a control system with relay With pulse frequency modulation PFM it is possible to achieve virt...

Page 94: ...able for your controller 1 7 Disturbance variable in the controller menu select the binary input to which you have connected the flow measured value Menu Additional functions Controllers Controller 1 Disturbance variable Source of data Binary input 6 2 and Measured value PFM value 8 You can activate a dosing pump through PFM via a digital output of the DIO module Connect the pump to an output of t...

Page 95: ...a a relay or a binary output for example 2 The cleaning trigger is transferred to another device via a binary output This does not have its own connected cleaning unit but its sensors are installed in the medium affected by cleaning the master and are set to hold by the trigger 3 The trigger is transferred via an additional binary output to another device whose connected sensors have their own cle...

Page 96: ...tion High Low Specifies the active signal level Low Input signals between 0 and 5 V DC High Input signals between 11 and 30 V DC 1 x y slot no input number 3 Link the binary inputs with the analyzer select Menu Setup Analyzer Extended setup Signal for process access 4 Select measuring channel SP1 or in the case of two channel devices select measuring channel SP1 or SP2 5 Assign a binary input to t...

Page 97: ... The signal is active for a configurable time before every activity that requires sample This makes it possible to activate an external pump or a dilution module for example 1 For the measurement under Menu Setup Analyzer Measurement select Continuous as the start condition analyzer measures continuously without interruption between the measurements 2 Select Menu Setup Inputs Binary input x y 3 Co...

Page 98: ...n external pump before every measurement If you convey the sample to the analyzer using an external pump or an external sample preparation system you can use the Signal for sample request to switch on the external devices for a limited time only when the analyzer requires sample The signal becomes active before every measurement calibration and cleaning You can configure how long the signal should...

Page 99: ...red 10 6 2 Binary input configuration Menu Setup Inputs Binary input x y 1 Function Options Info Binary input Selection Off On Factory setting On Switches the input on off Signal type Selection Static signal PFM Factory setting Static signal Select the signal type Static signal Use this setting to read out the position of for example an on off switch an inductive proximity switch or a PLC binary o...

Page 100: ...between 0 and 5 V DC High Input signals between 11 and 30 V DC Signal type PFM Max frequency 100 00 to 1000 00 Hz Factory setting 1000 00 Hz Maximum frequency of the PFM input signal Is to equal the maximum possible upper limit of the measuring range If the value selected is too small higher frequencies will not be detected If the value is too large on the other hand the resolution for small frequ...

Page 101: ...it of measure Input variable Parameter Customized text 16 characters Specify the unit for your parameter e g hPa Lower range value Input variable Parameter or Flow rate 2000 00 to 0 00 Factory setting 0 00 The start of measuring range corresponds to a frequency of 0 Hz Your previously defined unit is additionally displayed Upper range value Input variable Parameter or Flow rate 0 00 to 10000 00 Fa...

Page 102: ...ut on off Signal type Selection Static signal PFM Factory setting Static signal Select the signal type Static signal Comparable to a relay output of a diagnostic status a limit switch or the status of an active measurement PFM You can output a measured value e g the chlorine value or the manipulated variable of a controller It acts as a wear free switching contact that can be used to activate a do...

Page 103: ...ration of the limit switches Menu Setup Additional functions Limit switches Operating mode Function Diagnostics message Selection as assigned Namur M Namur S Namur C Namur F Factory setting as assigned as assigned With this selection the diagnosis messages are transmitted over the binary output that you individually allocated to it Namur M F When you choose one of the Namur classes all messages al...

Page 104: ...Operation Liquiline System CA80HA 104 Endress Hauser Menu Setup Outputs Binary output x y 1 Function Options Info Meas value format Selection Factory setting Specify the number of decimal places ...

Page 105: ...t should trigger the connected actuator e g the dosing pump Bipolar Split range Unipolar Part of the manipulated variable that the controller uses to increase the process value Unipolar For connected actuators that decrease the controlled variable Hold behavior Selection Freeze Fixed value None Factory setting None Freeze The device freezes the last value Fixed value You define a fixed current val...

Page 106: ... the sensors and controllers connected to inputs are available for selection Measured value Selection None Depends on the Source of data Factory setting None The measured value you can select depends on the option selected under Source of data The list of dependent measured values is provided in the Measured value table subject to the Source of data 107 Apart from measured values from connected se...

Page 107: ... 0 mA Factory setting 22 0 mA Specify which current should be output at this current output in the hold state 1 x y slot output number Measured value depending on the Source of data Source of data Measured value Total hardness Hardness Selection Main value pH Glass Selection Raw value mV pH Temperature pH ISFET ORP Selection Temperature ORP mV ORP Oxygen amp Selection Temperature Partial pressure ...

Page 108: ...roller 2 Temperature 1 3 Mathematical functions All the mathematical functions can also be used as a data source and the calculated value can be used as the measured value Outputting the controller manipulated variable via the current output Unipolar Assign to the output to which an actuator that can increase the measured value is connected Unipolar Assign to the output to which an actuator that c...

Page 109: ...ollowing functions can be output via a relay Limit switch status Controller manipulated variable for controlling an actuator Diagnostic messages Status of a cleaning function in order to control a pump or a valve You can assign a relay to multiple inputs in order to clean several sensors with just one cleaning unit for example Menu Setup Outputs Alarm relay or relay at channel no Function Options ...

Page 110: ...select all the limit switches in one go Hold behavior Selection Freeze last value Fixed value Ignore Factory setting Ignore Outputting the manipulated variable of a controller To output a controller manipulated variable via a relay the relay is modulated The relay is energized pulse t1 and is then de energized interval t0 Function Controller Function Options Info Source of data Selection None Cont...

Page 111: ...n remains constant 0 T t1 t0 2T 3T y t T 1 10 50 100 1 High 0 Low t 62 Typical application solenoid valve 2 PFM pulse frequency modulation Here pulses of a constant length t1 are output and the interval between the pulses varies t0 At a maximum frequency t1 t0 1 High 0 Low T t t1 t0 2T y f f T T max min 100 50 20 3T 63 Typical application dosing pump ...

Page 112: ...e Shortest turn on time Operating mode PWM 0 3 s to Cycle duration Factory setting 0 3 s Pulses that are shorter than this limit value are not output in order to conserve the actuator Maximum frequency Operating mode PFM 1 to 180 min 1 Factory setting 60 min 1 Maximum number of pulses per minute The controller calculates the pulse duration based on this setting Hold behavior Selection Freeze last ...

Page 113: ... M Namur S Namur C Namur F Factory setting Relay as assigned Alarm relays Namur F as assigned If this option is selected the diagnostic messages which you have individually assigned to the relay are output via the relay Namur M Namur F If you decided to use one of the Namur classes all the messages that are assigned to the individual class are output via the relay You can also change the Namur cla...

Page 114: ...er Cleaning type Chemoclean Plus 4x Cleaning 1 0V 4x Cleaning 2 0V 1 Hold behavior Selection Freeze last value Fixed value Ignore Factory setting Ignore Freeze last value The device freezes the last measured value Fixed value You define a fixed measured value that is output at the output Ignore A hold has no effect 1 0V is variable text which you can assign in Menu Setup Additional functions Clean...

Page 115: ...hose status is to be transmitted PROFIBUS PROFINET variables PROFIBUS PROFINET device A maximum of 4 analog AO and 8 digital DO PROFIBUS variables are available as measured values in the controller limit switch or current output menus Example Using an AO or DO value as the set point for the controller Menu Setup Additional functions Controller 1 1 In the specified menu define PROFIBUS as the data ...

Page 116: ...e to be output 3 Define how the device should behave in the hold state for sensors Configuration options of Source of data Measured value and Hold behavior 107 4 In the case of controllers also specify the type of manipulated variable Please note that if you select Hold behavior Freeze the system not only flags the status but also actually freezes the measured value In addition you can define 8 di...

Page 117: ...nctions controllers and sets for measuring range switching Measuring value Selection Depends on Source of data Select the measured value see the following table Measured value depending on the Source of data Source of data Measured value Total hardness Hardness Selection Main value pH Glass Selection Raw value mV pH Temperature pH ISFET ORP Selection Temperature ORP mV ORP Oxygen amp Selection Tem...

Page 118: ...3 NO3 N Ultrasonic interface Selection Interface SAC Selection Temperature SAC Transm Absorption COD BOD Controller 1 Current input 1 3 Selection Bipolar only for current outputs Unipolar Unipolar Controller 2 Temperature 1 3 Mathematical functions All the mathematical functions can also be used as a data source and the calculated value can be used as the measured value You can monitor the manipul...

Page 119: ...Limit value overshoot or undershoot 64 Measured value within or outside a range 65 Rate of change 67 Limit value Settings depend on the measured value Operating mode Above limit check or Below limit check 1 2 t MV t1 t2 t3 t4 t5 1 2 t MV t1 t2 t3 t4 t5 A B A0028523 64 Exceeding A and undershooting B a limit value without hysteresis and switch on delay 1 Limit value 2 Alarm range t1 3 5 No action t...

Page 120: ...rt delay Start delay has elapsed If measured values are decreasing the relay contact is reset when the switch off point is undershot Range upper value Hysteresis and after the drop out delay Switch off delay Hysteresis Settings depend on the measured value Operating mode Change rate 1 3 t MV t1 t2 2 A0028525 66 Hysteresis taking the example of limit value overshoot 1 Limit value 2 Alarm range 3 Hy...

Page 121: ... value Delta value an event is generated No further event is generated if the value continues to experience such a steep increase or decrease If the slope is back below the limit value the alarm status is reset after a preset time Auto Confirm Events are triggered by the following conditions in the example given t2 t1 Delta time and MV1 Delta value t4 t3 Auto Confirm and MV2 Delta value t6 t5 Delt...

Page 122: ... setting Off Switches the function on or off Start date 01 01 2000 to 31 12 2099 Format DD MM YYYY Enter start date Start time 00 00 00 to 23 59 59 Format hh mm ss Enter start time Signal duration 00 00 03 to 2400 00 00 Format hh mm ss Duration of the high signal level at the start of a cycle Period 00 00 03 to 2400 00 00 Format hh mm ss Duration of a cycle Signal Display only Current process valu...

Page 123: ...n Using the mathematical function an analog temperature set point is thus available for a controller T C t h 18 21 23 8 12 17 0 T2 T1 A0041704 69 Time based temperature control 1 Program Time switch 1 T1 Start date 01 01 2020 Start time 08 00 00 Signal duration 09 00 00 Period 24 00 00 2 Define Time switch 2 T2 Start date 01 01 2020 Start time 12 00 00 Signal duration 05 00 00 Period 24 00 00 3 Cr...

Page 124: ...ry 2 hours This only applies if the pH value is lower than 4 0 1 Program Time switch 1 Start date 01 01 2020 Start time 00 00 00 Signal duration 00 10 00 Period 02 00 00 2 Create Formula mathematical function Menu Setup Additional functions Mathematical functions MF1 Formula Tracking On Source A Time switch 1 Source B pH value of a Memosens pH input Formula A AND B 4 0 3 Use the formula as the dat...

Page 125: ...ed variable measured value X results in the control deviation which is filtered by a neutral zone The neutral zone is used to suppress control deviations e that are too small The control deviation thus filtered is now fed to the actual PID controller which divides into 3 parts based on the P proportional I integral and D derivative values top down The integral section middle also comprises an anti...

Page 126: ...ler Set point Setpoint Neutral zone Xn Proportional band Xp Integral action time I value Tn Derivative action time D value Tv 6 What should the controller do in the event of a hold measured error sensor replacement cleaning etc Pause or continue with dosing Hold behavior Manipulated Variable At the end of a hold continue or restart the control loop affects I value Hold behavior State 7 How should ...

Page 127: ...tpoint Current set point x Current measured value Disturbance variable Current measured value of disturbance variable Norm disturb value Name Free text Give the controller a name so you can identify it later on Controller Enable Selection None Binary inputs Limit switches Fieldbus variables Factory setting None In connection with the DIO module you can select a binary input signal e g from an indu...

Page 128: ...ime and the controller has to react to these changes on a continuous basis If the flow rate and medium remain constant the manipulated variable can also assume a fixed value once the process has settled Since the control process is never finished here this type of control is also referred to as continuous control A mixture of both process types can often be found in practice the semi batch process...

Page 129: ...f action the controller acts as a downwards controller It becomes active when the process value e g the temperature is too high range B y 100 0 100 w x Xp Xp A B 72 Red overlap between the curves of the two 1 sided controllers Controlled variable Source of data Selection None Sensor inputs Current inputs Fieldbus signals Binary inputs Mathematical functions Factory setting None Specify the input o...

Page 130: ...edforward control involves an open loop control system since its effect is not measured directly That means that the feed flow is directed exclusively forward Hence the designation feedforward control For the additive feedforward control that is alternatively available in the device the standardized disturbance variable is added to the controller manipulated variable This enables you to set up a t...

Page 131: ...urrent inputs Fieldbus signals Binary inputs Mathematical functions Factory setting None Specify the input which should be the source of data for the disturbance variable Measured value Selection Depends on Source of data Factory setting None Specify the measured value that should be your disturbance variable You can use different measured values depending on the data source Zp Range of adjustment...

Page 132: ...re the following Time constant Twu Time constant α Width of the neutral zone Xn Width of the hysteresis range of the neutral zone Xhyst Clock time of the controller Tn 0 0 to 9999 0 s Factory setting 0 0 s The integral action time specifies the effect of the I value If Tn 0 the following applies Clock Twu 0 5 Tn Tv The integral action time is the time needed in a step function response to achieve ...

Page 133: ...actory setting 0 3 Affects the additional damping filter of the D controller The time constant is calculated from α Tv Process balance Controller type PID 2 sided Selection Symmetric Asymmetric Factory setting Symmetric Symmetric There is only one control gain and this applies for both sides of the process Asymmetric You can set the control gain individually for both sides of the process Xp Proces...

Page 134: ...ting 0 0 Width of the hysteresis range of the neutral zone relative component of xn xn xhyst e w x y The graphic illustrates the manipulated variable with a pure P controller over the control deviation e set point minus controlled variable Very low control deviations are set to zero Control deviations xn are processed in the normal way Via xhyst it is possible to configure a hysteresis to suppress...

Page 135: ...one Factory setting All Select Should the hold trigger the previously selected exception behaviour or should it be ignored Outputs Goes to the menu Outputs Controller assignment view Shows an overview of the inputs and outputs used 10 8 4 Cleaning programs for sensors LCAUTION Cleaning not switched off during calibration or maintenance activities Risk of injury due to medium or cleaning agent If a...

Page 136: ... Cleaning Cleaning 1 4 Standard clean Function Options Info Cleaning time 5 to 600 s Factory setting 10 s Cleaning duration The cleaning duration and interval depend on the process and sensor Determine the variables empirically or based on experience Define the cleaning cycle 139 Chemoclean An example is the use of the CYR10 injector unit to clean pH glass sensors CYR10 connection 41 Menu Setup Ad...

Page 137: ...ut You can specify the number and name of the outputs further down in the menu See below for an example of programming Failsafe step setup Table view In the table specify the states that the relays or outputs should assume in the event of an error Limit contacts 0 to 2 Select the number of digital input signals e g from limit position switches of the retractable assembly Limit contact 1 2 Selectio...

Page 138: ...0 ES2 0 5 0 1 0 0 0 5 0 0 0 0 The pneumatic retractable assembly e g CPA87x is activated by compressed air via a two way valve As a result the assembly assumes either the Measure position sensor in medium or the Service position sensor in rinse chamber Media such as water or cleaning agents are supplied via valves or pumps Here there are two states 0 off or closed and 1 on or open The hardware req...

Page 139: ...00 00 The interval value can be between 1 minute and 7 days Example You have set the value 1 00 00 Each day the cleaning cycle starts at the same time you started the first cleaning cycle Daily event times Cleaning cycle Weekly program 00 00 to 23 59 HH MM 1 Define up to 6 times Event time 1 6 You can then choose from these for each day of the week 2 Choose individually for each day of the week wh...

Page 140: ...le Action Start an individual cleaning process with the selected parameters If cyclical cleaning is enabled there are times in which it is not possible to manually start the process Stop or Stop Failsafe Action End the cleaning process cyclically or manually Outputs Goes to the menu Outputs Cleaning program assignment view Shows an overview of the cleaning processes 10 8 5 Mathematical functions I...

Page 141: ...l function assignment view Overview of the configured functions Redundancy Use this function to monitor two or three sensors with redundant measurements The arithmetic average of the two closest measured values is calculated and output as the redundancy value Menu Setup Additional functions Mathematical functions MF1 to 8 Mode Redundancy Function Options Info Calculation Selection Off On Factory s...

Page 142: ...ensor Instead of the mathematical functions you can also connect a combined pH ORP sensor Set the main measured value simply to rH Menu Setup Additional functions Mathematical functions MF1 to 8 Mode rH calculation Function Options Info Calculation Selection Off On Factory setting Off On off switch for the function pH source Connected pH sensor Set the input for the pH sensor and the input for the...

Page 143: ...y calculating the degassed conductivity Menu Setup Additional functions Mathematical functions MF1 to 8 Mode Degassed conductivity Function Options Info Calculation Selection Off On Factory setting Off On off switch for the function Cation conductivity Connected conductivity sensor Cation conductivity represents the sensor downstream from the cation exchanger and upstream from the degassing module...

Page 144: ...end on the sensors connected Select the sensors that should function as the minuend Inlet e g sensor upstream from the ion exchanger or subtrahend Outlet e g sensor downstream from the ion exchanger Measured value Outlet Measured value Main value format Selection Auto Factory setting Auto Specify the number of decimal places Cond unit Selection Auto µS cm mS cm S cm µS m mS m S m Factory setting A...

Page 145: ...κv Inlet direct conductivity κh Outlet acid conductivity Inlet The options depend on the sensors connected Inlet Sensor upstream from the cation exchanger direct conductivity Outlet Sensor downstream from the cation exchanger acid conductivity Selection of the measured value is obsolete as it must always be Conductivity Measured value Outlet Measured value Calculated pH Read only View this value i...

Page 146: ...l EXP A LN Natural logarithm Numerical Numerical LN B LOG Decadic logarithm Numerical Numerical LOG A MAX Maximum of two values Numerical Numerical MAX A B MIN Minimum of two values Numerical Numerical MIN 20 B MOD Division with remainder Numerical Numerical MOD 10 3 ABS Absolute value Numerical Numerical ABS C NUM Boolean numerical conversion Boolean Numerical NUM A Equals Boolean Boolean A B Not...

Page 147: ... selected source are possible measured values 3 Enter the formula 4 Switch on the calculation The current measured values A B and C as well as the result of the calculation using the formula are displayed Measured value Selection Depends on the source A C The current measured value is displayed Formula User defined text Table 146 Make sure the exact notation is used upper case Blank spaces before ...

Page 148: ...cal function and connect the dosing pump to the corresponding relay The pump is switched on if all 3 conditions are met If one of the conditions is no longer met the pump is switched off again Instead of outputting the result of the formula directly to a relay you can also connect a limit switch in between in order to attenuate the output signal via a switch on and switch off delay Example Load ba...

Page 149: ...mit switch output enables the use of switch on switch off delay for example Menu Setup Additional functions Diagnostic modules Function Options Info Diagnostic module 1 961 8 968 Source of data Selection None Fieldbus signals Binary inputs Limit switches Factory setting None Specify the input which should be the source of data for the diagnostic message Measured value Selection depend on Source of...

Page 150: ...plausible the local display is faulty or you encounter other problems search for the faults under Process errors without messages Operating Instructions Memosens BA01245C or Device specific errors 154 Follow the recommended measures 4 Contact the Service Department if you cannot rectify the error yourself citing only the error number 11 1 2 Process errors without messages Operating Instructions Me...

Page 151: ...alibration does not start Action still active Appropriate bottles not used Check status No sample available Device is in the fieldbus mode no manual actions are possible Dispenser run times have elapsed Calibration has failed Stability criterion was not met 1 Check the configuration and perform manual calibration again 2 Check the hoses of the reagents Current output incorrect current value Incorr...

Page 152: ...mode C Function check no error Maintenance work is being performed on the device Wait until the work has been completed S Out of specification the measuring point is being operated outside its specification Operation is still possible However you run the risk of increased wear shorter operating life or lower measurement accuracy The cause of the problem is to be found outside the measuring point M...

Page 153: ...e message appears as Past message Possible settings The list of diagnostic messages displayed depends on the path selected There are device specific messages and messages that depend on what sensor is connected Menu Setup Run the Extended setup Diagnostics settings Diag behavior Function Options Info List of diagnostic messages Select the message to be changed Only then can you make the settings f...

Page 154: ... the message to an output Configure one of the output types mentioned as follows Menu Setup Outputs Alarm relay or Binary output or relay Function Diagnostics and Operating mode as assigned Cleaning program Selection None Cleaning 1 4 Factory setting None Decide whether the diagnostic message should trigger a cleaning program You can define cleaning programs under Menu Setup Additional functions C...

Page 155: ...ly 285 Update error F On On Firmware update failed 1 Repeat 2 SD card error use another card 3 Incorrect firmware repeat with suitable firmware 4 Contact the Service Department 302 Battery low M On Off Buffer battery of real time clock is low The date and time are lost if the power is interrupted Contact the Service Department battery replacement 304 Module data F On On At least 1 module has incor...

Page 156: ...temperature range not reached This could affect the functionality of the reagents 1 Check whether the insulation cap is seated correctly over the reagents 2 Replace cooling heating module 3 Contact the Service Department 362 Photometer temp F On Off Temperature of the photometer too high Contact the Service Department 363 Photometer temp F On Off Temperature of the photometer too low Contact the S...

Page 157: ...g system Contact the Service Department and quote the number displayed 381 Valve defective F On On Indicated valve is defective Contact the Service Department and quote the number displayed 385 Heating temp low S On Off Sample pre heating target temperature not reached 1 Check fuse 2 Check cable and connector 386 Heating temp high S On Off Temperature of sample pre heating too high 1 Monitor the t...

Page 158: ...ical function fault condition 1 Check mathematical function 2 Check assigned input variables 460 Output below limit S On Off Reasons Sensor in air Air pockets in assembly Sensor fouled Incorrect flow to sensor 1 Check sensor installation 2 Clean sensor 3 Change assignment of current outputs 461 Output above limit S On Off 502 No text catalog F On On Contact the Service Department 503 Language chan...

Page 159: ...ation and analyzers see the Operating Instructions for the sample preparation system 714 Filter mats change M On Off Filter mat change necessary The limit value for operating hours has been exceeded Replace the filter mats and reset the operating hours counter in the Diagnostics menu 715 Calibration expired M On Off The validity of the last calibration has expired The date of the last calibration ...

Page 160: ...ion 731 Leakage detected F On Off Leak detected in the sample preparation system or analyzer 1 Check hoses and connections 2 Check solenoid valves 3 Check leak sensor 4 Check the free outlet on the analyzer 5 Replace defective parts and if necessary reset the operating hours counters in the Diagnostics menu 732 Wear parts alarm F On On One or more wear parts has exceeded its service life Replace t...

Page 161: ... 951 958 Hold active CH1 C On Off Output values and status of the channels are on hold Wait until the hold is deactivated again 961 968 Diagnostic module 1 961 Diagnostic module 8 968 S Off Off Diagnostic module is enabled 969 Modbus Watchdog S Off Off The device did not receive a Modbus telegram from the master within the specified time The status of Modbus process values received is set to inval...

Page 162: ... CO2 conc range F On On CO2 concentration degassed conductivity outside the measuring range 992 pH calculation range F On On pH calculation outside the measuring range 993 rH calculation range F On On rH calculation outside the measuring range 994 Difference conduct F On On Dual conductivity outside the measuring range 1 Status signal 2 Diagnostic message 3 Failure current 11 6 2 Sensor specific d...

Page 163: ... Calibration events 75 Yes No Yes Yes Operation logbook Configuration events 250 Yes No Yes Yes Diagnostics logbook Diagnostic events 250 Yes No Yes Yes Analyzer event logbook Analyzer events 19500 2 No No Yes Yes Analyzer calibration logbook Analyzer calibration logbook 250 Yes No Yes Yes Analyzer data logbook Analyzer data logbooks 20000 2 No No Yes Yes Analyzer absorbance data logbook Absorptio...

Page 164: ...rmation The complete list is always visible however Delete all entries Action You can delete all the calibration logbook entries here Configuration events Chronological list of the configuration events Show Events are displayed Select a particular event to display more detailed information Go to date User entry Go to date Time Use this function to go directly to a specific time in the list In this...

Page 165: ... also adapt the display to suit your individual requirements Press the navigator button in the graphic display you are given additional options such as the zoom function and x y movement of the graph Define the cursor if you select this option you can move along the graph with the navigator and view the logbook entry data stamp measured value in text form for every point in the graph Simultaneous ...

Page 166: ...00 nA 34 00 nA 09 11 2011 14 02 20 23 11 2011 11 25 32 126 00 nA 34 00 nA 09 11 2011 14 02 20 23 11 2011 11 25 32 6 00 pH 09 11 2011 14 02 20 23 11 2011 11 25 32 09 11 2011 14 02 20 23 11 2011 11 25 32 12 00 pH A0016688 73 Simultaneous display of two graphs the top one is selected ...

Page 167: ...settings in the menu General settings Logbooks Delete all entries Action You can use this to delete all data logbook entries Absorption data logbook Curve Read only Displays the absorption curve Show Events are displayed Select a particular event to display more detailed information Go to date User entry Go to date Time Use this function to go directly to a specific time in the list In this way yo...

Page 168: ... data logbook entries for sensors Data logbook 1 8 Logbook name This submenu is available for each data logbook that you have set up and activated Source of data Read only Input or mathematical function is displayed Measured value Read only Measured value being recorded is displayed Log time left Read only Display of days hours and minutes until logbook is full Pay attention to the information on ...

Page 169: ... You can import the FDM files into Fieldcare and archive them so they are tamper proof The file name is made up of the Logbook ident Menu Setup General settings Logbooks an abbreviation for the particular logbook and a time stamp 1 CSV files use international number formats and separators Therefore they must be imported into MS Excel as external data with the correct format settings If you double ...

Page 170: ...ardware version Photometer Read only Firmware version Hardware version Modbus Only with the Modbus option Read only Enable Bus address Termination Modbus TCP Port 502 Modbus specific information PROFIBUS Only with the PROFIBUS option Read only Termination Bus address Ident number Baudrate DPV0 state DPV0 fault DPV0 master addr DPV0 WDT ms Module status and other PROFIBUS specific information Ether...

Page 171: ...s a time stamp The information is saved on the SD card in a sysinfo subfolder The csv file can be read and edited in MS Excel for example This file can be used when servicing the device 1 Provided you give the manufacturer all the information about changes to the hardware 11 11 Sensor information Select the channel you want from the list of channels Information in the following categories is displ...

Page 172: ...ch current output Simulation Selection Off On Factory setting Off If you simulate the value at the current output this is indicated on the display by a simulation icon in front of the current value Current 2 4 to 23 0 mA Factory setting 4 mA Set the desired simulation value Alarm relay Relay x y Simulation of a relay state This menu appears once for each relay Simulation Selection Off On Factory s...

Page 173: ...erature Selection Off On Factory setting Off If you simulate the temperature measured value this is indicated on the display by a simulation icon in front of the temperature Temperature 50 0 to 250 0 C 58 0 to 482 0 F Factory setting 20 0 C 68 0 F Set the desired simulation value Binary input x y Binary output x y Simulation of a binary input or output signal The number of submenus available corre...

Page 174: ...he settings Factory default Selection OK ESC Restart with factory settings Settings that have not been saved are lost 11 15 Operating time information List of functional and operating time counters Operating time filter mats Displays the period of usage in days Operating time photometer Remaining operating time Dispensers Displays the remaining period of usage in days i e the individual dispensers...

Page 175: ...ithout restriction including without limitation the rights to use copy modify merge publish distribute sublicense and or sell copies of the Software and to permit persons to whom the Software is furnished to do so subject to the following conditions The above copyright notice and this permission notice shall be included in all copies or substantial portions of the Software THE SOFTWARE IS PROVIDED...

Page 176: ...arge ESD Risk of damaging the electronic components Take personal protective measures to avoid ESD such as discharging beforehand at PE or permanent grounding with a wrist strap For your own safety use only genuine spare parts With genuine parts the function accuracy and reliability are also ensured after maintenance work NOTICE Failure to observe the maintenance intervals Possibility of material ...

Page 177: ...ce tasks 12 2 1 Software menu Maintenance Menu Operation Maintenance Bottle replacement If level monitoring is activated specify when bottles are replaced Dispenser replacement Operating time counter can be reset here Replace dispensers 183 Liquid manager replacement Operating time counter can be reset here Replace Liquid Manager 184 Filter mats replacement Operating time counter can be reset here...

Page 178: ...cleaning agents NOTICE Cleaning agents not permitted Damage to the housing surface or housing seal Never use concentrated mineral acids or alkaline solutions for cleaning Never use organic cleaners such as acetone benzyl alcohol methanol methylene chloride xylene or concentrated glycerol cleaner Never use high pressure steam for cleaning Cleaning CAT8x0 sample preparation automatic cleaning The Li...

Page 179: ...system with water 6 Trigger the action again Menu Operation Maintenance Decommissioning Rinse with water 7 Remove the hoses from the beaker and dry them with a clean paper towel Cleaning the optional sample collecting vessel Cleaning the collecting vessel Trigger the rinsing and draining of the collecting vessel before rinsing the analyzer 178 1 Remove all the hoses from the sample collecting vess...

Page 180: ...ove the bottles that you wish to replace by removing the hoses on the Liquid Manager Use a paper towel to catch any reagents that may escape 4 Select Menu Operation Maintenance Bottle change mode Bottle removal Bottle selection 5 Select the bottles that you want to remove and press OK to confirm 6 Select Confirm bottles removed Replacing bottles 1 Replace the removed bottles with fresh reagent Rea...

Page 181: ...he hose from the bottle Use a paper towel to catch any reagents that may escape This prevents any contamination of the zero standard 3 Replace the standard with zero standard 4 Connect the zero standard to the Liquid Manager connection S1 5 Select Menu Operation Maintenance Bottle change mode Bottle insertion Bottle selection 6 Select Standard S1 and press OK to confirm 7 Select Confirm bottles in...

Page 182: ...e should not be any more sample left in the sample collecting vessel 2 Remove the bottles and the bottle tray 3 Remove the cover of the carrier board 4 Replace the hoses with hoses of the same diameter and length Attach hose labels to the new hoses 5 Route drain hose D6 behind dispenser 7 6 Secure the cover and place the bottle tray back into the housing 7 Connect the bottles to the appropriate co...

Page 183: ...erating time filter mats Reset 12 2 8 Replacing dispenser You require the following parts From the CAV800 maintenance kit NORPRENE hose ID 1 6 mm 10 ml dispensers with adapter 2 5 ml dispenser with adapter 1 pair of gloves that are resistant to the reagents used Preparing to replace dispensers 1 Rinse the system 179 2 Remove the bottles including the reagent hoses and the bottle tray 3 Remove the ...

Page 184: ... Manager 5 Select Menu Operation Dispenser replacement Dispenser selection 6 Select the dispensers that you have replaced and press OK to confirm 7 Select Reset operating time counter 8 Select Menu Operation Maintenance Bottle change mode Bottle insertion Bottle selection 9 Select all the bottles and press OK to confirm 10 Select Confirm bottles inserted 11 The system must be calibrated after a re...

Page 185: ...d Manager 5 Install the new Liquid Manager in the reverse order Reinserting the dispensers 1 Screw a new dispenser onto the Liquid Manager Make sure that the dispenser is connected correctly 2 Slide the dispenser holder over the snap fit hooks Ensure that the holder snaps into place Reinserting the bottles and starting the analyzer again 1 Secure the cover and place the bottle tray back into the h...

Page 186: ...ervice as described To take out of service proceed as follows 1 Set the manual mode 2 Empty the sample collecting vessel 3 Rinse the system 12 3 1 Setting the manual mode Initial situation The device is still in the automatic mode and you now want to take the device out of service 1 Press the MODE soft key The current mode is displayed Automatic 2 Select an action Manual mode The device changes to...

Page 187: ...Maintenance Decommissioning Rinse analyzer once again to rinse the hoses with air 9 The analyzer can now be disconnected from the mains supply 13 Repair 13 1 General information The repair and conversion concept provides for the following The product has a modular design Spare parts are grouped into kits which include the associated kit instructions Only use original spare parts from the manufactu...

Page 188: ...ectronic equipment WEEE the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste Do not dispose of products bearing this marking as unsorted municipal waste Instead return them to Endress Hauser for disposal under the applicable conditions Dispose of batteries correctly Always dispose of batteries in accordance with local regulations on b...

Page 189: ...I01137C 07 EN The Liquiline System CAT860 can only be operated with a Liquiline System CA80 single channel device 14 1 2 Consumables You can find the order codes on the website https www endress com device viewer 1 Indicate the serial number of the device 2 Search Device information is displayed 3 Select the Spare parts tab 4 Click the product root The complete product structure is displayed The f...

Page 190: ...ion code Kit for upgrade with cooling system Cooling module integrated in base of housing Bottle tray with recess and insulation Activation code Kit for upgrade for second downstream analyzer 3 Valve for switching sample flow Hoses connection adapters Activation code Kit for dilution function Order No CAZ800 AAN5 Kit for upgrading the dilution function Hose with identification marking Modified cab...

Page 191: ...luted media With open aperture Digital with Memosens 2 0 technology Product Configurator on the product page www endress com cps91e Technical Information TI01497C Memosens CPS31E pH sensor for standard applications in drinking water and swimming pool water Digital with Memosens 2 0 technology Product Configurator on the product page www endress com cps31e Technical Information TI01574C Memosens CP...

Page 192: ... applications in process and environmental engineering Digital with Memosens 2 0 technology Product Configurator on the product page www endress com cps12e Technical Information TI01494C Memosens CPS42E ORP sensor for process technology Digital with Memosens 2 0 technology Product Configurator on the product page www endress com cps42e Technical Information TI01575C Memosens CPS72E ORP sensor for ...

Page 193: ...S21E Digital conductivity sensor for media with medium or high conductivity Conductive measurement With Memosens 2 0 Product Configurator on the product page www endress com cls21e Technical Information TI01528C Oxygen sensors Oxymax COS51D Amperometric sensor for dissolved oxygen With Memosens technology Product Configurator on the product page www endress com cos51d Technical Information TI00413...

Page 194: ...m cas40d Technical Information TI00491C Turbidity sensors Turbimax CUS51D For nephelometric measurements of turbidity and solids in wastewater 4 beam scattered light method With Memosens technology Product Configurator on the product page www endress com cus51d Technical Information TI00461C Turbimax CUS52D Hygienic Memosens sensor for turbidity measurement in drinking water process water and in u...

Page 195: ... Modbus RS485 71219871 Activation code for EtherNet IP 71135635 Activation code for Profibus DP for module 485 71449914 Upgrade code for EtherNet IP web server for BASE2 71449915 Upgrade code for Modbus TCP web server for BASE2 71449918 Upgrade code for web server for BASE2 71449901 Upgrade code for PROFINET web server for BASE2 71249548 Kit CA80 activation code for 1st digital sensor input 712495...

Page 196: ...Product Configurator on the product page www endress com cyz71d Technical Information TI00502C Field Data Manager Software MS20 21 PC software for central data management Visualization of series of measurements and logbook events SQL database for secure data storage 14 3 System components 14 3 1 Measuring cables Memosens data cable CYK10 For digital sensors with Memosens technology Product Configu...

Page 197: ...uts for sensors with Memosens protocol optional Analog current inputs optional Binary inputs optional 15 1 4 Input signal Depending on version 2 x 0 4 to 20 mA optional passive potentially isolated 15 1 5 Current input passive Span 0 to 20 mA Signal characteristic Linear Internal resistance Non linear Test voltage 500 V 15 1 6 Cable specification for optional sensors with Memosens technology Cable...

Page 198: ...d Galvanic isolation Yes Connectors Spring terminal max 1 5 mm bridged internally T function optional M12 Bus termination Internal slide switch with LED display Modbus RS485 Signal encoding EIA TIA 485 Data transmission rate 2 400 4 800 9 600 19 200 38 400 57 600 and 115 200 baud Galvanic isolation Yes Bus termination Internal slide switch with LED display Web server and Modbus TCP Signal encoding...

Page 199: ...Signal on alarm Adjustable as per NAMUR Recommendation NE 43 In measuring range 0 to 20 mA Failure current from 0 to 23 mA In measuring range 4 to 20 mA Failure current from 2 4 to 23 mA Factory setting for failure current for both measuring ranges 21 5 mA 15 2 3 Load Max 500 Ω 15 2 4 Transmission behavior Linear 15 3 Current outputs active 15 3 1 Span 0 to 23 mA 15 3 2 Signal characteristic Linea...

Page 200: ...elays 2 0 A Relay switching capacity Base module Alarm relay Switching voltage Load max Switching cycles min 230 V AC cosΦ 0 8 to 1 0 1 A 700 000 0 5 A 450 000 115 V AC cosΦ 0 8 to 1 0 1 A 1 000 000 0 5 A 650 000 24 V DC L R 0 to 1 ms 0 1 A 500 000 0 5 A 350 000 Extension module Switching voltage Load max Switching cycles min 230 V AC cosΦ 0 8 to 1 0 1 A 700 000 0 5 A 450 000 2 A 120 000 115 V AC ...

Page 201: ...tion cyclical communication master class 1 to slave 1 MSAC1 connection acyclical communication master class 1 to slave 2 MSAC2 connections acyclical communication master class 2 to slave Device lock The device can be locked using the hardware or software Addressing using DIL switches or software GSD PDM DD DTM 15 5 2 Modbus RS485 Protocol RTU ASCII Function codes 03 04 06 08 16 23 Broadcast suppor...

Page 202: ... software 15 5 4 EtherNet IP Log EtherNet IP ODVA certification Yes Device profile Generic device product type 0x2B Manufacturer ID 0x049Eh Device type ID 0x109F Polarity Auto MIDI X Connections CIP 12 I O 6 Explicit message 6 Multicast 3 consumers Minimum RPI 100 ms default Maximum RPI 10000 ms System integration EtherNet IP EDS Rockwell Add on Profile Level 3 Faceplate for Factory Talk SE IO dat...

Page 203: ...h Device description files GSD Information and files under www endress com On the product page for the device Documents Software Device drivers www profibus com On the website under Products Product Finder Polarity Auto polarity for automatic correction of crossed TxD and RxD pairs Supported connections 1 x AR IO Controller AR 1 x AR IO Supervisor Device AR connection allowed 1 x Input CR Communic...

Page 204: ...nfiguration Factory setting 15 5 6 Web server The Web server enables full access to the device configuration measured values diagnostic messages logbooks and service data via standard WiFi WLAN LAN GSM or 3G routers with a user defined IP address TCP port 80 Supported features Remote controlled device configuration 1 session Save restore device configuration via SD card Logbook export file formats...

Page 205: ...ightened with 2 Nm 15 6 6 Sensor connection optional Sensors with Memosens protocol Sensor types Sensor cable Sensors Digital sensors without additional internal power supply With plug in connection and inductive signal transmission pH sensors ORP sensors Combined sensors Oxygen sensors amperometric and optical Conductivity sensors with conductive measurement of conductivity Chlorine sensors disin...

Page 206: ... Additional measured error depending on the temperature 1 5 μA K 15 7 4 Repeatability 5 0 3 mg l ppm With dilution and dilution factor 5 Additionally 1 5 of end of measuring range 15 7 5 Repeatability of sensor inputs Documentation of the connected sensor 15 7 6 Measuring interval Continuous approx 7 min adjustable 10 min to 24 h 15 7 7 Sample requirement Without dilution module 14 7 ml 0 50 fl oz...

Page 207: ...ication 15 7 12 Maintenance effort Weekly visual inspection Quarterly 1 hour 15 8 Environment 15 8 1 Ambient temperature range 5 to 40 C 41 to 104 F 15 8 2 Storage temperature 20 to 60 C 4 to 140 F 15 8 3 Humidity 10 to 95 non condensing 15 8 4 Degree of protection IP55 cabinet analyzer stand TYPE 3R cabinet analyzer stand 15 8 5 Electromagnetic compatibility 6 Interference emission and interferen...

Page 208: ...rsion Weight with cooling module Weight without cooling module Cabinet version 42 kg 92 6 lbs 39 5 kg 87 1 lbs Open installation 34 kg 74 96 lbs 31 5 kg 69 45 lbs Analyzer stand 75 kg 165 3 lbs 72 5 kg 159 8 lbs 15 10 3 Materials Parts not in contact with medium Cabinet version exterior cover Plastic ASA PC Open installation exterior cover Cabinet version interior lining Plastic PP Open installati...

Page 209: ...ic PMMA Plastic PP Stainless steel 1 4404 V4A EPDM Valve optional PVDF 15 10 4 Process connection Sample inlet With sample collecting vessel Plug in connector for rigid hoses with OD 4 mm Without sample collecting vessel Hose barb for flexible hoses with ID 1 6 mm Dilution water Hose barb for flexible hoses with ID 3 2 mm Outlet Hose barb for flexible hoses with ID 13 mm 15 10 5 Hose entries 4 x b...

Page 210: ...ting 24 Basic settings 62 Basic setup 59 Bus address 44 Bus termination 43 C Cable entries 205 Cable specification 197 205 Calculated pH value 145 Changing the password 79 Check Connection 45 Installation 25 Installation and function 58 Chemoclean 41 136 Chemoclean Plus 137 Chlorine regulation with feedforward control 93 Cleaning cycle 139 Cleaning function 114 Cleaning programs Chemoclean 136 Che...

Page 211: ...ment architecture 15 EtherNet IP 50 74 116 202 Event logbook 163 Exporting the setup 77 Extended setup 71 F Fieldbus Connecting 42 Termination 43 Firmware history 175 Firmware update 76 Formula 145 Function check 58 H Hardware settings 43 Hold settings 64 Hook and loop Velcro fasteners 195 Hose connection diagram 55 Hose entries 209 Hose specification 209 Humidity 207 I Incoming acceptance 16 Inpu...

Page 212: ...rating the measuring device into the system 46 Modbus 115 Relay 109 Signal outputs 106 Operational safety 7 Output Output signal 198 Relay outputs 200 Outputs Binary 92 Current outputs 106 EtherNet IP 116 Modbus 115 PROFIBUS DP 115 PROFINET 115 Relay 109 Signal outputs 106 P Pollution degree 208 Port assignment 15 Power consumption 204 Power supply 204 24 V version 29 Cable entries 205 Cable speci...

Page 213: ...y 109 Signal outputs 106 Web server 74 Signal outputs 106 Simulation 172 Slot assignment 15 Soft keys in the measuring mode 60 Spare parts 187 Standard cleaning 136 Standard solution 189 State of the art technology 8 Storage temperature 207 Supply voltage 204 Switching on 59 Symbols 5 System information 169 System integration Fieldbus 49 Service interface 48 Web server 46 System test 173 T Technic...

Page 214: ...www addresses endress com 71532810 71532810 ...

Reviews: