Liquiline System CA80HA
Operation
Hauser
71
4. Only in the case of
Fill up buffer
:
Decide to set
Overflow warning
:
On
or
Off
.
On
: The device displays a diagnostic message in the event of memory overflow.
5.
Line plotter
submenu: Specify the type of graphic representation.
10.2.5 Extended setup
Diagnostics settings
The list of diagnostic messages displayed depends on the path selected. There are device-
specific messages, and messages that depend on what sensor is connected.
Menu/Setup/(General settings or Inputs<sensor channel>)/Extended setup/Diagnostics settings/Diag. behavior
Function
Options
Info
List of diagnostic messages
‣
Select the message to be changed. Only
then can you make the settings for this
message.
Diag. code
Read only
Diagnostic message
Selection
• On
• Off
Factory setting
Depends on the message
You can deactivate or reactivate a diagnostic
message here.
Deactivating means:
• No error message in the measuring mode
• No failure current at the current output
Failure current
Selection
• On
• Off
Factory setting
Depends on the message
‣
Decide whether an failure current should
be output at the current output if the
diagnostic message display is activated.
In the event of general device errors, the
failure current is output at all the
current outputs. In the event of channel-
specific errors, the failure current is only
output at the assigned current output.
Status signal
Selection
• Maintenance (M)
• Out of specification (S)
• Function check (C)
• Failure (F)
Factory setting
Depends on the message
The messages are divided into different error
categories in accordance with NAMUR NE 107.
‣
Decide whether you want to change a
status signal assignment for your
application.