Liquiline System CA80HA
Maintenance
Hauser
181
4. Select all the bottles that you have replaced and press
OK
to confirm.
5. Select
Confirm bottles inserted
.
6. If bottle level monitoring is activated (
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Diagnostics settings
/
Bottles
), you must reset this under
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Reset filling levels
.
7. Following the replacement, a zero point calibration must be performed and the
calibration factor must be redetermined:
Menu
/
Operation
/
Manual operation
/
Determine zero point
and /
Determine calibration factor
.
8. Go back to the
MODE
/
Continue automatic mode
or
MODE
/
Start automatic mode
to
start the normal measurement.
12.2.5 Carrying out zero point calibration
Replacing standard calibration solution with zero standard
1. If you have not already done so,
set the manual mode.
2. Remove the bottles containing standard S1 by removing the hose from the bottle. Use a
paper towel to catch any reagents that may escape. This prevents any contamination of
the zero standard.
3. Replace the standard with zero standard.
4. Connect the zero standard to the Liquid Manager, connection S1.
5. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle
selection
.
6. Select
Standard S1
and press
OK
to confirm.
7. Select
Confirm bottles inserted
.
Determining the zero point
1. Call up:
Menu
/
Operation
/
Manual operation
/
Determine zero point
.
2. Carry out zero point calibration.
3. Following successful calibration, the following message appears: "Do you want to accept
the calibration data for adjustment?". Press
OK
to confirm.
Re-deploying the standard calibration solution
1. Replace the zero standard with the standard. Use either an additional hose in order to
reconnect the standard calibration solution, or allow the hose to drain and dry the hose
well.
2. Connect the standard calibration solution to the Liquid Manager.
3. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle
selection
.
4. Select
Standard S1
and select
OK
to confirm.