Operation
Liquiline System CA80HA
114
Hauser
Outputting the status of a cleaning function
Function = Cleaning(for sensors)
Function
Options
Info
Assignments
Selection
• None
• Depends on the type of
cleaning
Factory setting
None
Here you can specify how a cleaning function
should be displayed for the relay.
You have the following options depending on
the cleaning program that is selected (
Menu
/
Setup
/
Additional functions
/
Cleaning
:
•
Cleaning type
=
Standard clean
Cleaning 1 - Water
,
Cleaning 2 - Water
,
Cleaning 3 - Water
,
Cleaning 4 - Water
•
Cleaning type
=
Chemoclean
Cleaning 1 - Water
,
Cleaning 1 - Cleaner
,
Cleaning 2 - Water
,
Cleaning 2 - Cleaner
,
Cleaning 3 - Water
,
Cleaning 3 - Cleaner
,
Cleaning 4 - Water
,
Cleaning 4 - Cleaner
•
Cleaning type
=
Chemoclean Plus
4x
Cleaning 1 - %0V
, 4x
Cleaning 2 -
%0V
Hold behavior
Selection
• Freeze last value
• Fixed value
• Ignore
Factory setting
Ignore
Freeze last value
The device freezes the last measured value.
Fixed value
You define a fixed measured value that is
output at the output.
Ignore
A hold has no effect.
1)
%0V is variable text which you can assign in
Menu
/
Setup
/
Additional functions
/
Cleaning
/
Chemoclean Plus
/
Output label 1 ... 4
.
Outputting the "Measurement active" system status and the information "Sample
required"
Function = Analyzer
Function
Options
Info
Signal type
=
Static signal
Assignments
Function
=
Analyzer
Selection
• None
• Sample required SP1
• Measurement active SP1
Factory setting
None
If this option is selected, information is output
as to whether a measurement is active on the
selected measuring channel or an action is
started that requires sample (measurement,
calibration or cleaning).