Liquiline System CA80HA
Operation
Hauser
137
Chemoclean Plus
An example is the use of the CYR10 injector unit to clean pH glass sensors. (CYR10 connection
Menu/Setup/Additional functions/Cleaning/Cleaning 1 ... 4/Chemoclean Plus/ChemoCleanPlus setup
Function
Options
Info
Cleaning steps setup
Table to create a time program You define a maximum of 30 program steps
which should be performed one after the other.
For each step enter the duration [s] and the
state (0 = "off", 1 = "on") of each relay or output.
You can specify the number and name of the
outputs further down in the menu.
See below for an example of programming.
Failsafe step setup
Table view
‣
In the table specify the states that the
relays or outputs should assume in the
event of an error.
Limit contacts
0 to 2
‣
Select the number of digital input signals
(e. g. from limit position switches of the
retractable assembly).
Limit contact 1 ... 2
Selection
• Binary inputs
• Fieldbus signals
‣
Define the signal source for each limit
position switch.
Outputs
0 to 4
‣
Select the number of outputs that
actuators, such as valves or pumps, should
activate.
Output label 1 ... 4
Free text
You can assign a meaningful name to each
output, e. g. "assembly", "cleaner 1", "cleaner 2",
etc.