Menu/Setup/Inputs/Binary input x:y
Function
Options
Info
Input variable
Selection
• Frequency
• Parameter
• Flow rate
Factory setting
Frequency
Frequency
Display in Hz in the measuring menu
Parameter
You subsequently determine the parameter
name and unit. These are then displayed in the
measuring menu.
Flow rate
For connecting a flowmeter
Parameter name
Input variable
=
Parameter
Customized text, 16 characters
‣
Define a name for the parameter, e.g.
"Pressure".
Unit of measure
Input variable
=
Parameter
Customized text, 16 characters
‣
Specify the unit for your parameter, e.g.
"hPa".
Lower range value
Input variable
=
Parameter
or
Flow rate
-2000.00 to 0.00
Factory setting
0.00
The start of measuring range corresponds to a
frequency of 0 Hz.
Your previously defined unit is additionally
displayed.
Upper range value
Input variable
=
Parameter
or
Flow rate
0.00 to 10000.00
Factory setting
0.00
The end of measuring range corresponds to the
maximum frequency defined above.
Your previously defined unit is additionally
displayed.
Damping
0 to 60 s
Factory setting
0 s
The damping causes a floating average curve of
the measured values over the time specified.
1)
x:y = slot no. : input number