Operation
Liquiline System CA80HA
144
Hauser
Dual conductivity
You can subtract two conductivity values and use the result to monitor the efficiency of an ion
exchanger, for example.
Menu/Setup/Additional functions/Mathematical functions/MF1 to 8/Mode = Dual conductivity
Function
Options
Info
Calculation
Selection
• Off
• On
Factory setting
Off
On/off switch for the function
Inlet
The options depend on the
sensors connected
Select the sensors that should function as the
minuend (
Inlet
, e.g. sensor upstream from the
ion exchanger) or subtrahend (
Outlet
, e.g.
sensor downstream from the ion exchanger).
Measured value
Outlet
Measured value
Main value format
Selection
• Auto
• #
• #.#
• #.##
• #.###
Factory setting
Auto
Specify the number of decimal places.
Cond. unit
Selection
• Auto
• µS/cm
• mS/cm
• S/cm
• µS/m
• mS/m
• S/m
Factory setting
Auto
Dual conductivity
Read only
View this value in a user-defined measuring
screen or output the value via the current
output.
Mathematical function
assignment view
Overview of the configured functions