Liquiline System CA80HA
Operation
Hauser
103
Menu/Setup/Outputs/Binary output x:y
Function
Options
Info
Signal type
=
Static signal
Function
Selection
• None
• Limit switches
• Diagnostics message
• Cleaning
• Analyzer
Factory setting
None
Source for the outputted switching state
The following functions depend on the option
selected.
Function
=
None
switches off the function.
There are no other settings.
Assignments
Function
=
Cleaning
Multiple selection
•
Cleaning 1 - Water
...
• Cleaning 4 - Cleaner
Here you can decide which binary outputs
should be used for activating the valves and
pumps. Here you specifically assign a control
signal to the binary output for the cleaner/
water dosing of a cleaning program.
You can define cleaning programs under:
Menu
/
Setup
/
Additional functions
/
Cleaning
.
Data sources
Function
=
Limit switches
Multiple selection
Limit switch 1 ... 8
‣
Select the limit switches that should be
output via the binary output.
Configuration of the limit switches:
Menu
/
Setup
/
Additional functions
/
Limit switches
.
Operating mode
Function
=
Diagnostics message
Selection
• as assigned
• Namur M
• Namur S
• Namur C
• Namur F
Factory setting
as assigned
as assigned
With this selection, the diagnosis messages are
transmitted over the binary output that you
individually allocated to it.
Namur M ... F
When you choose one of the Namur classes, all
messages allocated to that respective class are
output.
You can change the Namur class assignment
for every diagnostic message.
Assignments
Function
=
Analyzer
Selection
• None
• Sample required SP1
• Measurement active SP1
Factory setting
None
If this option is selected, the binary output
outputs information as to whether a
measurement is active on the selected
measuring channel or an action is started that
requires sample (measurement, calibration or
cleaning).
In the case of two-channel devices,
Sample
required SP2
and
Measurement active SP2
is
also displayed
Signal type
=
PFM
Max. frequency
1.00 to 1000.00 Hz
Factory setting
1000.00 Hz
Maximum frequency of the PFM output signal
Is to equal the maximum possible upper limit
of the measuring range.