Maintenance
Liquiline System CA80HA
186
Hauser
12.3 Taking out of service
L
CAUTION
Activities while the analyzer is in operation
Risk of injury and infection from medium!
‣
Before you release any hoses, make sure that no actions, such as the pumping of sample,
are currently running or are due to start shortly.
‣
Wear protective clothing, goggles and gloves or take other suitable measures to protect
yourself.
‣
Wipe up any spilt reagent with a disposable tissue and rinse with clear water. Then dry the
cleaned areas with a cloth.
NOTICE
Do not operate the analyzer for longer than 5 days without taking it out of service
properly!
Can result in damage to the device.
‣
Take the analyzer out of service as described.
To take out of service, proceed as follows:
1. Set the manual mode
2. Empty the sample collecting vessel
3. Rinse the system
12.3.1 Setting the manual mode
Initial situation: The device is still in the automatic mode and you now want to take the device
out of service.
1. Press the
MODE
soft key.
The current mode is displayed:
Automatic
2. Select an action:
Manual mode
The device changes to the manual mode.
3. Wait until all operations have stopped.
4.
Alternatively:
Stop actions currently running under:
Menu
/
Operation
/
Manual operation
.
12.3.2 Emptying the sample collecting vessel
1.
Menu
/
Operation
/
Maintenance
/
Decommissioning
/
Sample collector
/
Empty sample
collector
/
Start
must be selected.
The system is automatically rinsed with sample and emptied. This process takes
approx. 4 minutes.
2. In the case of a two-channel device, repeat the procedure for the second sample
collecting vessel.