CMP 200
®
X-Ray Generator Service Manual Ch # 901476-10 Rev. C
Page 8-7
8.2.0
SPARE PARTS LIST (Cont)
2.
The AEC board in your generator was selected to be compatible with specific AEC devices. To
maintain full compatibility, the original part number must be ordered as a replacement. Refer to
chapter 9, section 9.2.0, for the part number of the original AEC board that was shipped in the
generator for which this manual was prepared.
3.
This applies to the dual speed starter option only, spares should be stocked accordingly. The part
number shown for the dual speed starter board is for the board only, without the phase-shift
capacitors. For the complete dual speed starter subassembly, several part numbers (which are tube-
stator dependent) are used in CMP 200
®
DR 400 / 480 VAC generators. To determine which dual
speed starter subassembly is in your generator, note the DUAL SPEED STARTER ASSY part number
on a label near the top left side of the dual speed starter chassis. This will be the part number that
must be ordered for spares usage. For reference, this will be part number 901297-XX or 901298-XX
where XX is a two digit number designating the exact configuration.
4.
This generator control board is jumper configurable. Please refer to the following tables to configure
the generator control board for your x-ray generator.
WARNING:
DO NOT ATTEMPT TO RECONFIGURE 125 kV MAXIMUM GENERATORS FOR 150 kV
OPERATION. SERIOUS GENERATOR DAMAGE WILL OCCUR.
TABLE 8-5
FUNCTION JUMPER
CONFIGURATION
Firmware
Upgrading
JW2
PINS 1-2: Normal
PINS 2-3: Upgrade
JW4 and JW5
Set as per Table 8-5b below.
kV Overvoltage
Trip
JW6
PINS: 1-2: 125 kV
PINS: 2-3: 150 kV
JW7
Not used.
Cooling Fan
JW12
PINS 1-2: Fan
PINS 2-3: No Fan
Maximum kV
JW13
PINS 1-2: 125 kV (For 125 kV tank)
PINS 2-3: 150 kV (For 150 kV tank)
DAP JW14
Set as per
DAP Installation
in Chapter 3C.
HT Tank
Feedback Comp.
JW15
PINS 1-2: 125 kV
PINS 2-3: 150 kV
TABLE 8-5b
32 / 40 kW
400 / 480 VAC
125 kV
50 kW
400 / 480 VAC
150 kV
32 / 40 kW
208 / 230 VAC
125 kV
32 / 40 kW
400 / 480 VAC
150 kV
50 kW
208 / 230 VAC
150 kV
65 kW
400 / 480 VAC
150 kV
JW4
Pins 1-2
Pins 2-3
Pins 2-3 Pins
1-2 Pins
2-3 Pins
2-3
JW5
Pins 1-2
Pins 1-2
Pins 2-3 Pins
1-2 Pins
2-3 Pins
2-3
5.
These assemblies apply to generators equipped with a membrane console only.