CPI Canada Inc.
Interfacing and Programming
3C
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-49
3C.7.3 Receptor Setup (Cont)
FUNCTION
(MEMBRANE
CONSOLE)
FUNCTION
(GenWare
®
)
DESCRIPTION
THE FOLLOWING SELECTIONS ARE ONLY AVAILABLE IF DEFAULTS WAS ENABLED AS PREVIOUSLY DESCRIBED.
TECHNIQUE Technique
Defines which technique will be defaulted to when a receptor is selected.
Membrane console:
MA:
Defaults to mA/ms mode.
MAS:
Defaults to mAs mode.
AEC
Defaults to AEC mode.
GenWare
®
:
ms:
Defaults to mA/ms mode.
mAs:
Defaults to mAs mode.
AEC
Defaults to AEC mode.
FOCUS Focus
Defines which focus will be defaulted to when a receptor is selected. Options are
SMALL
or
LARGE
.
FILM SCREEN
Film Screen
Defines which film screen will be defaulted to when a receptor is selected and AEC enabled.
Options are film screen
1
,
2
, or
3
(membrane console) or
I
,
II
, or
III
GenWare
®
.
LEFT FIELD
CENTER FIELD
RIGHT FIELD
Fields
Defines which field(s) will be defaulted to when a receptor is selected.
Membrane console:
YES:
The selected field will be selected.
NO:
The selected field will not be selected.
GenWare
®
:
Refer to the graphic under
Fields:
A black (blue for Touchscreen GenWare
®
) field select rectangle = field not selected, a green
field select rectangle = field selected.
KV Voltage
(kV)
Selects the default kV.
MA Current
(mA)
Selects the default mA.