CPI Canada Inc.
Troubleshooting
5
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-07 Rev. B
Page 5-10
5.2.3
Error Messages (Cont)
ERROR
CODE
MESSAGE
(membrane console)
MESSAGE
(touchscreen console)
PROBLEM
ACTION
E065
TOMO DEVICE ERR.
Tomo device error.
The tomo start signal was not
received within 30 seconds of
pressing the X-ray switch on the
generator.
1. Check the tomo
connections to the
generator.
2. Check the tomo system.
E067
PS DUTY LIMIT
Power supply duty cycle
limit exceeded
Power supply duty cycle limit
exceeded.
Re-evaluate technique factors.
Allow generator to cool if
possible. Exposures are
inhibited until the generator
has cooled sufficiently.
E071
DAP DOSE OVERFLW
DAP overflow error
The accumulated DAP value
exceeds the display limit.
Reset the DAP.
E072
DAP DEVICE ERR.
DAP device error
The DAP device is not functional.
1. Check the DAP wiring.
2. Check the DAP interface
board.
E073
DAP DATA ERROR
DAP data error
The DAP configuration data is
corrupted.
Reset factory defaults.
E074
INVERTER 1 ERROR
Inverter 1 Error
Power circuit failure.
Contact product support.
E075
INVERTER 2 ERROR
Inverter 2 Error
Power circuit failure.
Contact product support.
E077
RES. CIRCUIT ERR
Resonant Circuit Error
HT primary overcurrent detected.
1. Check for arcing of the X-
ray tube.
2. Check for arcing in the HT
tank; if failure of HT tank is
suspected, contact product
support.
E078 BUCKY1
INTERLOCK
Bucky 1 Interlock error
Indicates that Bucky 1 is not ready.
Check Bucky 1.
E079 BUCKY2
INTERLOCK
Bucky 2 Interlock error
Indicates that Bucky 2 is not ready.
Check Bucky 2.
E080
INTERLOCK 1 OPEN
Interlock 1 Open
Indicates that the interlock 1 is open.
Check the circuits connected
to interlock 1. Refer to
INPUTS
in chapter 3C.
E081
INTERLOCK 2 OPEN
Interlock 2 Open
Indicates that the interlock 2 is open.
Check the circuits connected
to interlock 2. Refer to
INPUTS
in chapter 3C.