CMP 200
®
X-Ray Generator Service Manual Ch # 901476-01 Rev. C
Page 1-19
1.3.2
Generator Power Requirements (Cont)
Generator Series
and Mains Voltage
Minimum
Recommended
Mains
Disconnect to
Generator
(15 ft/5 m max)
Generator
Momentary
Line Current
Minimum
Recommended
Generator
Service Rating
Minimum
Recommended
Distribution
Transformer
Rating
Minimum
Recommended
Ground
Wire
Size
*
Apparent
Mains
Resistance
32 kW
208 VAC, 1p.
#4
(21 mm
2
)
220 A
120 A
45 kVa
#4
(21 mm
2
)
0.06
Ω
32 kW
230 VAC, 1p.
#4
(21 mm
2
)
200 A
120 A
50 kVa
#4
(21 mm
2
)
0.06
Ω
32 kW
208 VAC, 3p.
#4
(21 mm
2
)
120 A
100 A
45 kVa
#4
(21 mm
2
)
0.09
Ω
32 kW
230 VAC, 3p.
#4
(21 mm
2
)
110 A
100 A
45 kVa
#4
(21 mm
2
)
0.09
Ω
32 kW
400 VAC, 3p.
#6
(13.3 mm
2
)
65 A
100 A
45 kVa
#6
(13.3 mm
2
)
0.27
Ω
32 kW
480 VAC, 3p.
#6
(13.3 mm
2
)
55 A
100 A
45 kVa
#6
(13.3 mm
2
)
0.40
Ω
40 kW
208 VAC, 1p.
#2
(33 mm
2
)
275 A
120 A
65 kVa
#2
(33 mm
2
)
0.045
Ω
40 kW
230 VAC, 1p
#2
(33 mm
2
)
250 A
120 A
65 kVa
#2
(33 mm
2
)
0.045
Ω
40 kW
208 VAC, 3p.
#4
(21 mm
2
)
150 A
100 A
55 kVa
#4
(21 mm
2
)
0.075
Ω
40 kW
230 VAC, 3p.
#4
(21 mm
2
)
135 A
100 A
55 kVa
#4
(21 mm
2
)
0.075
Ω
40 kW
400 VAC, 3p.
#6
(13.3 mm
2
)
80 A
100 A
45 kVa
#6
(13.3 mm
2
)
0.22
Ω
40 kW
480 VAC, 3p.
#6
(13.3 mm
2
)
65 A
100 A
45 kVa
#6
(13.3 mm
2
)
0.32
Ω
50 kW
208 VAC, 3p.
#4
(21 mm
2
)
160 A
100 A
65 kVa
#4
(21 mm
2
)
0.055
Ω
50 kW
230 VAC, 3p.
#4
(21 mm
2
)
145 A
100 A
65 kVa
#4
(21 mm
2
)
0.055
Ω
50 kW
400 VAC, 3p
#6
(13.3 mm
2
)
100 A
100 A
65 kVa
#6
(13.3 mm
2
)
0.17
Ω
50 kW
480 VAC, 3p.
#6
(13.3 mm
2
)
80 A
100 A
65 kVa
#6
(13.3 mm
2
)
0.24
Ω
65 kW
400 VAC, 3p
#6
(13.3 mm
2
)
125 A
100 A
85 kVa
#6
(13.3 mm
2
)
0.13
Ω
65 kW
480 VAC, 3p
#6
(13.3 mm
2
)
105 A
100 A
85 kVa
#6
(13.3 mm
2
)
0.19
Ω
80 kW
400 VAC, 3p
#6
(13.3 mm
2
)
155A 100A 105
kVa
#6
(13.3 mm
2
)
0.10
Ω
80 kW
480 VAC, 3p
#6
(13.3 mm
2
)
130A 100A 105
kVa
#6
(13.3 mm
2
)
0.15
Ω
*
Refer to 1.3.3 for general grounding information.