CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-119
3C.13.10 Touchscreen Console Firmware Upgrade / Restore (Cont)
Figure 3C-43: Checking for Available Updates.
Step Action
9.
A list of current and available upgrade packets will be listed on the touchscreen console in the
Checking for Available Updates
screen. Select the packets you wish to install by touching the box next to the packet such that a check is placed inside the box
(as shown in figure 3C-43). Please note that packets that require updating will already be selected. Press
NEXT
.
Note: If no upgrade is available for the U-boot packet, it will not be available for selection. This is done to prevent
software corruption should the touchscreen console inadvertently lose power.
NOTE: ANY UPGRADE PACKET THAT IS OF AN EARLIER VERSION OF THE ONE ALREADY INSTALLED WILL APPEAR
IN RED.