2
Installation
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 2-18
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-02 Rev. C
2.7.2
Low Speed Starter Tube Select Table (Cont)
TABLE 2-1: SUPPORTED TUBE TYPES (LOW SPEED STARTER)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
BOOST
VOLTAGE
SHIFT CAPAC
H.V. AUXILIARY
BOARD PART NO.
Varian
B100
DX52
Std “R” stator
A102
A132
A142
240 VAC
33
μ
F
739442-00
739445-00
Varian
B100
“STD” stator
A102
A132
A142
240 VAC
33
μ
F
739442-00
739445-00
Varian
B130
B150
Std “R” stator
A192
A272
A282
A286
A292
G256
G292
240 VAC
33
μ
F
739442-00
739445-00
Varian
Diamond
Std “R” stator
RAD13
RAD14 0.3/1.2
RAD14 0.6/1.2
RAD14 0.6/1.5
240 VAC
33
μ
F
739442-00
739445-00
Varian
Emerald
Std “R” stator
RAD 8
RAD 68
RAD 74
240 VAC
33
μ
F
739442-00
739445-00
Varian
Saphire
Std “R” stator
RAD21
RAD56 0.6/1.0
RAD56 0.6/1.2
RAD60
RAD92
RAD94
240 VAC
33
μ
F
739442-00
739445-00
Varian DX62
300-400 kHu,
“STD” stator
A192B
A197
A256
A272
A282
A286
A292
240 VAC
33
μ
F
739442-00
739445-00
Varian DX62U
Universal
300-400 kHu,
configured as “STD” or
“R” stator
A192B
A197
A256
A272
A282
A286
A292
240 VAC
33
μ
F
739442-00
739445-00