CPI Canada Inc.
Regular Maintenance
6
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-08 Rev. B
Page 6-7
6.4.0
OIL FILL / LEVEL CHECK (Cont)
Figure 6-1: HT tank oil fill
6.5.0 CLEANING
•
Never use anything other than mild soap and water to clean plastic surfaces. Other cleaners may damage
the plastic.
•
Never use any corrosive, solvent or abrasive detergents or polishes.
•
Ensure that no water or other liquid can enter any equipment. This precaution prevents short circuits and
corrosion forming on components.
•
Methods of disinfection used must conform to legal regulations and guidelines regarding disinfection and
explosion protection.
•
If disinfectants are used which form explosive mixtures of gases, these gases must have dissipated before
switching on the equipment again.
•
Disinfection by spraying is not recommended because the disinfectant may enter the X-ray equipment.
•
If room disinfection is done with an atomizer, it is recommended that the equipment be switched OFF,
allowed to cool down and covered with a plastic sheet. When the disinfectant mist has subsided, the
plastic sheet may be removed and the equipment be disinfected by wiping.