3C
Interfacing and Programming
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-84
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.13.1 Accessing the Utilities Menu
Use these steps to access the systems utilities functions.
Step Action
1.
From the main console menu (figure 3C-22), press
System Utilities
. A pop-up window will be displayed (figure 3C-23),
requesting a password.
2.
Press
1
,
9
,
7
,
3
in sequence
to continue. This is the factory-default service password, and allows access to all of the functions
listed above.
•
Press
Clear
to cancel an incorrect password.
•
Press
Cancel
to return to the main menu.
•
Press
Accept
to access the system utilities menu. After a brief delay, the system utilities menu (figure 3C-24) will be
displayed. The message
Access Denied
indicates that an incorrect password was used.
The factory-default password
may be changed by a service engineer as described later in this supplement. If this was done, the password defined
above will not allow access to the system utilities menu.
Figure 3C-22: Main menu