2
Installation
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 2-8
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-02 Rev. C
2.6.1
Control Console (Cont)
Figure 2-3: Console cable routing
DO NOT CONNECT UNAPPROVED EQUIPMENT TO THE REAR OF THE CONSOLE.
For the membrane console, J3 is for connection of an external hand switch, J4 is a serial port
for use by an external computer, and J8 is for the interconnect cable to the generator main
cabinet.
For the touchscreen console, GEN on the rear of the touchscreen is for the interconnect
cable to the generator, HS is for connection of an external hand switch, COM 1 & COM 2 are
serial ports for use by external devices, LO (3.5 mm stereo jack) is for customer supplied
speakers (minimum 8 ohms, do not use externally amplified speakers), ETH is a standard
10/100 Ethernet connection, USBA and USBB are USB ports for connection of external
devices, CF is for the compact flash memory card which holds the touchscreen software, and
SW1 is the console upgrade button.
For the mini-console, the short piece of console cable with the attached 15-pin “D” connector
is for connection of the console cable that is connected to J19 on the generator control
board.
INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY RESULT IN
INJURY OR EQUIPMENT DAMAGE
.