X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.5.3 Statistics (Cont)
Use these steps to view the generator statistics.
Step
Action (membrane console)
Action (PC GenWare
®
)
Action (Touchscreen GenWare
®
)
1.
From the
UTILITY
menu, select
STATISTICS
.
Select
Generator Statistics
from the
Utility
menu, or use the statistics button
on the GenWare
®
toolbar.
Use the statistics button
on the
GenWare
®
toolbar to access the
Generator Statistics
window.
2.
The
STATISTICS
menu displays
exposure data, and allows the exposure
counter to be reset as described below:
•
The
TUBE 1 EXP
counter displays
the tube 1 exposure count made
since this counter was last reset.
•
TOTAL EXP
displays the total rad
exposure count. This is not resettable
*
•
Select
RESET TUBE 1 EXP
to reset
the tube 1 exposure counter.
The
Generator Statistics
window
displays exposure data, and allows the
exposure counter to be reset as
described below:
•
The
Tube 1 Rad Exposures
counter
displays the tube 1 exposure count
made since this counter was last
reset.
•
Total Number of Rad Exposures
displays the total rad exposure count.
This is not resettable
*
.
•
Select
Reset Tube 1
to reset the tube
1 exposure counter.
The
Generator Statistics
window
displays exposure data, and allows the
exposure counter to be reset as
described below:
•
The
Tube 1 Rad Exposures
counter
display the tube 1 exposure count
made since this counter was last
reset.
•
Total Number of Radiographic
Exposures
displays the total rad
exposure count. This is not resettable
*
.
•
Select
Reset Tube 1
to reset the tube
1 exposure counter.
*
The subject exposure counter is reset when the factory defaults are reset. Therefore, the “total exposure” count should be
recorded before resetting the factory defaults.