3C
Interfacing and Programming
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-52
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.7.3 Receptor Setup (Cont)
Step
Action (membrane console)
Action (PC GenWare
®
)
Action (Touchscreen GenWare
®
)
10.
Press
>>
.
11.
Select
SF/LF SWITCH
. Toggle the button
to select
AUTO
or
MAN
.
This function is not available in GenWare
®
12.
Select the
AEC
tab.
Select the
AEC
tab.
13.
Select
AEC BACKUP
. Toggle the button
to select
FIXED,
MAS,
or
MS
.
Under
AEC Mode
, select
fixed, mAs
, or
ms
.
Under
AEC Back-up Mode
, select
Fixed, mAs
, or
ms
.
14.
Select
AEC BACKUP MAS
. Use the
+
or
–
buttons to select the maximum backup
mAs.
Select the maximum backup mAs via the
AEC Back-Up mAs
dialog box.
Select the maximum backup mAs via the
AEC Backup mAs
dialog box.
15.
Select
AEC BACKUP MS
. Use the
+
or
–
buttons to select the maximum backup
ms.
Select the maximum backup ms via the
AEC Back-Up ms
dialog box.
Select the maximum backup ms via the
AEC Backup ms
dialog box.
16.
Select
AEC CHANNEL
. Use the
+
or
–
buttons to assign the desired AEC
channel to the selected receptor, or to
disable AEC operation on that receptor.
Select the AEC channel to be assigned to
the selected receptor, or disable AEC
operation on that receptor via the
AEC
Channel
dialog box.
Select the AEC channel to be assigned
to the selected receptor, or disable AEC
operation on that receptor via the
AEC
Channel
dialog box.
17.
Press
>>
. Select
the
Receptor Properties
tab.
Select the
Receptor Properties
tab.
18.
Select
RAD HANG
. Use the
+
or
–
buttons to select the desired rad
hangover time.
Select the desired rad hangover time via
the
Rad Hangover (s)
dialog box.
Select the desired rad hangover time via
the
Rad Hangover (s)
dialog box.
The following steps only apply if
MEMORY
in step 6 was set to
DEF
/
default
.
19.
Select
DEFAULTS
. Select
the
Receptor Defaults
tab.
Select the
Receptor Defaults
tab.
Select
TECHNIQUE
. Toggle the button to
select
MA,
MAS,
or
AEC
.
Under
Technique
, select
ms, mAs
, or
AEC
.
Under
Technique
, select
ms, mAs
, or
AEC
.
20.
Select
FOCUS
. Toggle the button to
select
SMALL
or
LARGE
.
Under
Focus
, select
small
or
large
. Under
Focus
, select
Small
or
Large.
Micro
focus is not available at this time.
21.
Select
FILM SCREEN
. Toggle the button
to select
1,
2,
or
3.
Under
Film Screen
, select
I, II,
or
III
. Under
Film Screen
, select
I, II,
or
III
.