CPI Canada Inc.
Interfacing and Programming
3C
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-51
3C.7.3 Receptor Setup (Cont)
NOTE:
DO NOT SWITCH OFF THE GENERATOR WHILE IN
*RECEPTOR SETUP DEFAULTS
* MENUS 4 AND 5. DOING SO WILL
CAUSE THE UPDATED RECEPTOR SETUP PARAMETERS NOT TO BE SAVED. IT IS RECOMMENDED THAT THE FIRST
RECEPTOR PROGRAMMING BE COMPLETED, THE RECEPTOR SETUP MENUS BE EXITED TO THE GEN CONFIGURATION
MENU, THEN THE RECEPTOR SETUP MENU BE RESELECTED TO PROGRAM THE NEXT RECEPTOR. THE ABOVE SHOULD
BE REPEATED UNTIL ALL RECEPTORS ARE PROGRAMMED. THIS WILL ENSURE THAT THE UPDATED PARAMETERS ARE
SAVED.
Use these steps to set up the receptor parameters. Refer to the definitions in the previous table.
Step
Action (membrane console)
Action (PC GenWare
®
)
Action (Touchscreen GenWare
®
)
1.
From the
GEN CONFIGURATION
menu,
select
RECEPTOR SETUP
.
Select
Receptor Setup
from the
Setup
menu, or use the receptor setup button
on the GenWare
®
toolbar.
Press the
button on the
GenWare
®
toolbar to access the
Receptor Setup
utility.
2.
Select the
Receptor Properties
tab.
Select the
Receptor Properties
tab.
3.
Select the first receptor to be
programmed.
Select the first receptor to be
programmed.
Select the first receptor to be
programmed.
4.
Select
TUBE
. Toggle the button to select
NONE
or
1
.
Check the
Receptor Enable
checkbox to
enable the selected receptor.
Check the
Receptor Enable
checkbox
to enable the selected receptor.
5.
Select
RECEPTOR SYM
. Use the
+
or
–
buttons to select the desired receptor
symbol.
6.
Select
MEMORY
. Toggle the button to
select
NO,
YES,
or
DEF
.
Under
Memory
, select
off, on
, or
default
. Under
Memory
, select
off, on
, or
default
.
7.
Select
TOMO
. Toggle the button to select
YES
or
NO
.
Check the
Tomo
checkbox to enable
tomographic operation.
Check the
Tomo
checkbox to enable
tomographic operation.
8.
Select
TOMO BUT
. Use the
+
or
–
buttons to select the desired tomo backup
time.
Select the desired tomo backup time via
the
Tomo Back-Up Time (ms)
dialog
box.
Select the desired tomo backup time via
the
Tomo Back-Up Time (ms)
dialog
box.
9.
Select
INTERFACE OPTS:
Use the
+
or
–
buttons to select the desired interface
option.
Select the desired interface option via the
Interface Options
dialog box.
Select the desired interface option via
the
Interface Options
dialog box.