3C
Interfacing and Programming
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-100
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.13.6 Touchscreen Setup (Cont)
Figure 3C-30: Touchscreen setup window, Settings tab
•
The
System Code
is a unique code assigned to each touchscreen on which the software is installed.
•
The
License
is a code specifically assigned to each touchscreen. This enables options in the features tab. Please consult the factory to
obtain a new license code if it is desired to enable new features, or to reinstall the existing license should the license code be accidentally
overwritten.
•
The
Customer Code
is a unique code assigned to each customer and selects the color scheme and graphic style (skin) to be displayed
on the touchscreen console.