CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-37
3C.7.1 Tube Selection / Tube Setup (Cont)
Step
Action (membrane console)
Action (PC GenWare
®
)
Action (Touchscreen GenWare
®
)
12.
Select
FIL BOOST
. Use the
+
or
–
buttons to set the filament rapid boost
duration.
Set the filament rapid boost duration via
the
Filament Boost Time (ms)
dialog
box.
Set the filament rapid boost duration via
the
Filament Boost Time (ms)
dialog
box.
13.
Select
FIL PREHEAT
. Use the
+
or
–
buttons to set the filament preheat time.
Set the filament preheat time via the
Filament Preheat Time (ms)
dialog box.
Set the filament preheat time via the
Filament Preheat Time (ms)
dialog box.
14.
Press
>>
.
Select the
Tube Settings
tab.
Select the
Tube Settings
tab.
15.
Select
MAX SF MA
. Use the
+
or
–
buttons to set the maximum small focus
mA.
Set the maximum small focus mA via the
Max. SF Current (mA)
dialog box.
Set the maximum small focus mA via the
Max. SF Current (mA)
dialog box.
16.
Select the
Anode Heating
tab.
17.
Select
ANODE HU WARNING
. Use the
+
or
–
buttons to set desired anode HU
warning %.
Set the desired anode HU warning % via
the
Anode Warning (%)
dialog box.
Set the desired anode HU warning % via
the
Anode Warning (%)
dialog box.
18.
Select
ANODE HU LIMIT
. Use the
+
or
–
buttons to set desired anode HU limit %.
Set the desired anode HU limit % via the
Anode Limit (%)
dialog box.
Set the desired anode HU limit % via the
Anode Limit (%)
dialog box.
19.
Press
Return
or alternatively
<<
to exit to
the
Tube 1 Selection
screen.
20.
Press
EXIT
as required to return to the
GEN CONFIGURATION
menu.
Typically, the boost time should be between 200 and 250 msec, and the preheat time should be in the range of 700 - 800 ms.
If in doubt, monitor the filament feedback and be sure that there is no change in the signal level 5 ms. after the start of an exposure, and that
the mA starts at the selected level.
Standby current must be below the emission point.
If the maximum filament current is increased, be careful not to exceed the tube manufacturer’s specifications.