3C.13.5 Receptor
Symbols
The current image receptor symbols may be replaced with predefined symbols chosen from the receptor symbols library or downloaded from
an external device.
Each image receptor button will always select a predefined image receptor (i.e. table Bucky, wall Bucky, DR, etc.). Before changing
the image receptor symbols, it must be clearly understood which image receptors are selected by each image receptor button. Each image
receptor button should then have a logical and intuitive symbol assigned to that position.
Use these steps to change the receptor symbols.
Step Action
1.
From the system utilities menu, press
Receptor Symbols
.
2.
A pop-up window will display (figure 3C-29) showing the image receptor buttons with the currently assigned symbols near the
right side of the receptor symbols window, and the library of available receptor symbols near the left side of the window.
3.
Select the image receptor button for which the symbol is to be changed. The receptor numbers in figure 3C-29 correspond to the
receptor numbers in GenWare
®
, and are shown for reference only. For the CMP 200
®
series, only receptors 1 to 5 are available.
4.
Select an appropriate symbol for the selected receptor from the symbols library. The selected symbol will be highlighted.
5.
Press
to assign the selected symbol to the selected image receptor button.
6.
Repeat steps 3 to 5 for each image receptor whose symbol is to be changed.
7.
Press
OK
to continue or
CANCEL
to return to the system utilities menu without making any changes. If
OK
was pressed, a pop-
up window will display asking if you wish to save the changes. Press
YES
to save the changes. Press
NO
to return to the system
utilities menu;
CANCEL
cancels the changes.
8.
To delete unused symbols from the receptor symbols library, select the symbol to be deleted. The selected symbol will be
highlighted.
•
Press
.
•
A pop-up window will display asking if you wish to delete the selected item. Press
YES
to delete the item;
NO
cancels the
deletion.
•
DO NOT DELETE SYMBOLS YOU MAY WANT TO USE IN THE FUTURE.
9.
To copy receptor symbol files to and from a USB flash drive press
to access the
USB File Transfer
Utility