CPI Canada Inc.
Troubleshooting
5
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-07 Rev. B
Page 5-6
5.2.3
Error Messages (Cont)
ERROR
CODE
MESSAGE
(membrane console)
MESSAGE
(touchscreen console)
PROBLEM
ACTION
E008
KV/MA FAULT
kV / mA Fault
The generator has detected a fault in
the kV or mA output during an
exposure and immediately
terminated the exposure. This may
be caused by arcing in the X-ray
tube, arcing of the HV cables, or HT
tank.
1. If arcing of the X-ray tube is
suspected, check condition
of tube. The X-ray tube
may be damaged or simply
require “seasoning”. Refer
to chapter 6 for tube
seasoning procedure.
2. If failure of HT tank is
suspected, contact product
support.
E009
PS NOT READY
Power Supply Not Ready
The generator is not ready to make
an exposure.
Retry exposure.
E011
HIGH MA FAULT
mA During Exposure
Too High
Generator CPU detected mA greater
than allowed tolerance.
Recalibrate X-ray tube
. *
E012
LOW MA FAULT
mA During Exposure
Too Low
Generator CPU detected mA less
than allowed tolerance.
Recalibrate X-ray tube
. *
E013
MANUAL TERMIN.
Manually Terminated
Exposure
Operator released exposure switch
during exposure.
1. Re-take exposure if
necessary.
2. Check for faulty switch
contacts or wiring.
E014
AEC BUT ERROR
AEC Back-up Timer -
Exposure Terminated
AEC exposure exceeded allowed
back up time.
1. Check exposure technique
settings.
2. Check that correct AEC
chamber is energized.
E015
AEC BU MAS ERR.
AEC mAs Exceeded -
Exposure Terminated
AEC exposure exceeded allowed
back up mAs.
1. Check exposure technique
settings.
2. Check that correct AEC
chamber is energized.
E016
TOMO BUT ERROR
Tomo Back-up Timer –
Exposure Terminated
Tomo exposure exceeded backup
time.
1. Check exposure technique
settings.
2. Increase tomo backup time,
if necessary.