1
Pre-installation
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-16
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-01 Rev. C
1.2.3
Safety Warning Labels (Cont)
DANGEROUS
VOLTAGE
High voltage (approximately 325 to 670 VDC, depending on model) is present on the inverter
assembly and associated components whenever the AC mains is energized and the console
is switched on, and for up to 5 minutes after the console is switched off or the AC mains is
disconnected.
THIS COMBINATION OF HIGH VOLTAGE AND HIGH CURRENT IS
POTENTIALLY LETHAL. USE EXTREME CAUTION WHEN SERVICING THIS UNIT.
1.3.0
PREPARING FOR INSTALLATION
1.3.1
Generator Heat Output
The maximum heat output of the main generator cabinet is less than 1000 BTU / hour in normal clinical
use, with a maximum of 70 BTU / hour heat output for the console. The console is convection-cooled, and
the main cabinet is fan cooled. The console and main cabinet should never be covered when the
generator is switched on, as any covering may interfere with the cooling.
1.3.2
Generator Power Requirements
The tables in this section show mains power requirements for various configurations of CMP 200
®
and
CMP 200
®
DR X-ray generators. The installer must ensure that the generator is connected to the proper
mains voltage as per the nameplate on the generator.
The table below depicts the power requirements for the 32 kW CMP 200
®
/ CMP 200
®
DR X-ray
generators.
Line Voltage
208 VAC - 5% to 230 VAC + 10%, 1 phase.
208 VAC - 5% to 230 VAC + 10%, 3 phase.
400 VAC
±
10%, 3 phase.
480 VAC
±
10%, 3 phase.
Line Frequency
50/60 Hz.
Momentary Current
200 Amps at 230 VAC (1 phase).
110 Amps/phase at 230 VAC (3 phase).
65 Amps/phase at 400 VAC.
55 Amps/phase at 480 VAC.
Nominal Current
*
≤
5 Amps.
Momentary Power Consumption
45 kVA.