Calibration Procedures
5-12
Service Manual
Exhalation PSOL Regulator
1.
Remove the orange tubing from the IN exhalation PSOL valve port and connect to a calibrated high
pressure gauge (see
2.
Using a flat head screwdriver, adjust the exhalation PSOL regulator, which is located near the exhalation
PSOL valve, to between 15.0 psig and 15.5 psig.
3.
Reconnect the orange tubing to the IN exhalation PSOL valve port.
Figure 5-4.
Exhalation PSOL/Safety Regulator
Safety Regulator
1.
While in diagnostics mode d3, adjust
Flow
to 3 L/min.
2.
Disconnect the patient circuit from the exhalation valve, and connect it to the Y-piece of the patient
circuit. This creates a closed loop.
3.
Tee in a low-pressure gauge in the patient circuit.
4.
Using a flat head screwdriver, adjust the safety regulator (see
) to obtain a
machine 1
pressure
reading of 140 ±5 cmH
2
O on the screen.
1
Orange tubing to exhalation
PSOL valve IN port
3
Exhalation PSOL regulator
2
Exhalation PSOL valve
4
Safety regulator
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......