Software Upgrade Procedure
Service Manual
2-29
Figure 2-25.
Circuit Check Screen (Step 2)
11.
When the circuit test is completed successfully, the message “Circuit Check PASSED” is displayed.
12.
If the circuit test failed, the message “Circuit Check FAILED” is displayed. Check the breathing circuit for
leaks, and repeat the circuit check. If the circuit check fails again, contact the Technical Service
Department for further assistance.
13.
Press the
Setup & Calibration
button on the control panel.
14.
Touch the
Technical
button on the screen (see
).
Figure 2-26.
Technical Screen
15.
Touch the
Regional Settings
16.
Touch the
Altitude
button on the screen. The “_ ft/_ m” setting will start flashing.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......