Procedures
Service Manual
5-5
Figure 5-2.
Analog PCB
1.
Enter diagnostics mode by performing the steps in section
, “
.” “d0” is displayed in the
Trig
window on the control panel and “0” is displayed in the
Flow/t Insp
window.
2.
Press the button below the
Trig
window.The display will blink.
3.
While the display is blinking, turn the
Adjustment
knob to change the setting from “d0” to “d5”.
4.
To confirm the selection, press the
Accept
button.
Note:
Steps 5 to 15 are to calibrate zero pressure offset of the exhalation drive pressure transducer (XD107).
Note:
Reference
for tubing, transducer, and zero pressure offset values.