![Covidien Newport e360 Service Manual Download Page 48](http://html1.mh-extra.com/html/covidien/newport-e360/newport-e360_service-manual_2673907048.webp)
Maintenance, Overhaul, and Software Upgrade
2-24
Service Manual
3.
When the
Logo
screen is displayed, release the
Accept
button.
4.
Wait for a few minutes until the
Software Download/Diagnostics Mode
screen is displayed (see
Figure 2-19.
Software Download / Diagnostics Mode Screen
5.
Connect the USB flash drive (with latest software) to the USB port on the back panel of the ventilator
(see
6.
Touch
Start Download
on the
Software Download/Diagnostics Mode
screen (
). The ventilator
will sound a short beep and the message “Touch Selection to Begin” changes to a flashing message
stating “downloading….” Depending on the software version your ventilator is currently running, the
download may take anywhere from 2 minutes to 45 minutes.
7.
When the software download is complete, the ventilator will sound a short beep. The message
“Download Complete” is displayed.
8.
Turn the ventilator off.
9.
Wait at least 10 seconds, and then turn the power back on.
10.
When the message “Ventilation Standby” appears, touch
Start Ventilating
to begin ventilating.
Note:
Because air/O
2
gas supplies are not connected, the Air/O
2
Loss alarm will sound. Disregard the alarm while
verifying the software is installed correctly.
11.
Confirm the message “Incompatible Software ….” does not appear on the screen. If this message
appears on the screen, the software download was NOT completed successfully. Repeat steps 1–6
again. If the download is unsuccessful again, contact the Technical Service Department for further
assistance.
12.
Press the
Extended Functions
button on the front control panel.
13.
Touch the
Event History
button on the screen.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......