Procedures
Service Manual
5-11
Figure 5-3.
Air/Oxygen Pressure Regulators Adjustment
14.
Adjust
Flow
as shown in
, and verify that the expected regulated pressure for each flow setting
is within tolerance.
Oxygen Inlet Regulator
1.
While in diagnostic mode d3, adjust FiO
2
to 100% (1.00) oxygen.
2.
Adjust
Flow
to 3 L/min.
3.
Adjust the oxygen inlet regulator pressure to 15.5 ±0.5 psig by following steps 11 through 13 in the
section titled “
.”
4.
Adjust
Flow
as shown in
, and verify that the expected regulated pressure for each flow setting
is within tolerance.
5.
Adjust
Flow
to 0 L/min and FiO
2
to 21% (0.21) once the calibration is complete.
1
Adjustment bolt
2
Lock nut
Table 5-4.
Air Flow and Pressure Settings
Flow setting
Air inlet pressure
3 L/min
15.0-16.0 psig
100 L/min
12.0-14.5 psig
180 L/min
>9.5 psig
Table 5-5.
Oxygen Flow and Pressure Settings
Flow setting
Oxygen inlet pressure
3 L/min
15.0-16.0 psig
100 L/min
12.0-14.5 psig
180 L/min
>9.5 psig
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......