2-1
2 Maintenance, Overhaul, and Software Upgrade
2.1
Overview
This chapter describes preventive maintenance, overhaul, and software upgrade procedures for
use on the Newport™ e360 ventilator.
2.2
Maintenance and Overhaul Intervals
The level I preventive maintenance procedure should be performed once a year or every 5000
hours, whichever comes first.
Perform the level 2 overhaul procedure every five years or 25 000 hours, whichever comes first.
Replace the lead acid battery and the O
2
sensor at least every 2 years, or when not performing as
intended.
2.3
General Warnings
WARNING:
To prevent personal injury or death, do not attempt any ventilator service or enter service mode while
a patient is connected to the ventilator.
WARNING:
Hazardous voltages are present inside the ventilator. Use extreme caution if it is necessary to work
inside the ventilator while it is connected to a power source. Disconnect electrical power, air, and
oxygen sources before attempting any disassembly. Failure to do so could result in injury to service
personnel or damage to equipment.
WARNING:
To maintain grounding integrity, the ventilator must be connected to a hospital-grade receptacle
when in use.
WARNING:
There is a risk of explosion if the ventilator is used in the presence of flammable anesthetics.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......