Operational Verification
6-6
Service Manual
6.5.2
Circuit Leak Test
Figure 6-3.
Circuit Leak Test
Note:
The circuit check test can only be performed in ventilation standby.
1.
Remove test lung, if present, and occlude the end of the patient circuit Y-piece using CAP100P.
2.
Turn on the ventilator.
3.
Touch the
Circuit Check
button on the screen, and follow the on-screen instructions.
4.
After the test is completed, verify that the message “PASSED” is displayed. Document the results on the
record sheet.
5.
Reattach the test lung.
6.
To exit ventilation standby, touch the
Start Ventilating
button on the screen.
6.5.3
Exhalation Flow Sensor and FiO
2
Sensor Calibration
1.
Touch the
Sensors
button on the screen.
2.
Touch the
Flow Sensor
button on the screen, and follow the on-screen instructions.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......