Operational Checks
Service Manual
6-13
8.
Repeat step 7 with
Tidal Volume
settings of 500 ml and 1 L.
6.5.16
Expiratory Tidal Volume
1.
If using the TSI certifier, connect the calibration analyzer at the patient wye on the patient circuit. If using
a PTS2000, connect the calibration analyzer on the expiratory side of the breathing circuit.
2.
Set
Tidal Volume
to 250 mL.
3.
Verify that the value of VTE displayed on the
Numeric
screen is between 225-275 mL, and that the tidal
volume reading measured by the calibration analyzer and the
Tidal Volume
setting are within ±10% of
each other. Document the results on the record sheet.
4.
Set
Tidal Volume
to 500 mL.
5.
Verify that the value of VTE displayed on the
Numeric
screen is between 450-550 mL, and that the tidal
volume reading measured by the calibration analyzer and the
Tidal Volume
setting are within ±10% of
each other. Document the results on the record sheet.
6.
Set
Tidal Volume
to 1.00 L.
7.
Verify that the value of VTE displayed on the
Numeric
screen is between 900 mL to 1.1 L, and that the tidal
volume reading measured by the calibration analyzer and the
Tidal Volume
setting are within ±10% of
each other. Document the results on the record sheet.
6.5.17
Audio Paused
1.
Create an alarm condition by changing the settings.
2.
Wait for the alarm to sound.
3.
Press the
Audio Paused
button, and verify the indicator lights up and the audible alarm stops.
4.
Press the
Audio Paused
button again, and verify the indicator turns off and the audible alarm resumes.
5.
Return the ventilator to standard test settings.
6.
To clear alarm indicators, press the
Reset
button.
7.
Document the results on the record sheet.
6.5.18
Loss of Power Alarm and Battery Operation
1.
Disconnect the power cord from the AC power source.
2.
Verify the following occurs:
•
The messages “AC Power Loss” and “Battery Back Up” are displayed in the message window
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......