Operational Verification
6-10
Service Manual
6.
Verify that the value of t1 (t Insp) displayed on the
Numeric
screen, the inspiratory time reading measured
by the calibration analyzer, and the t lnsp setting are within ±0.05 seconds of each other. Document the
results on the record sheet.
7.
Set
t lnsp
to 0.50 seconds.
8.
Verify that the value of t1 (t Insp) displayed on the
Numeric
screen, the inspiratory time reading measured
by the calibration analyzer, and the t lnsp setting are within ±0.10 seconds of each other. Document the
results on the record sheet.
6.5.9
Respiratory Rate
1.
Set the calibration analyzer to breath rate measurement (with 10 slpm).
2.
Set
t lnsp
to 0.25 seconds.
3.
Attach the test lung to the patient circuit Y-piece.
4.
Set
Resp Rate
to 5 b/min.
5.
Verify the value of RRtot displayed on the calibration analyzer and the
Resp Rate
setting are within ±1
b/min. Document the results on the record sheet.
6.
Repeat step 4 with a
Resp Rate
setting of 20 b/min.
7.
Set
Resp Rate
to 10 b/min.
6.5.10
Pressure Control
1.
Set the ventilator to pressure control.
2.
Set the calibration analyzer to peak pressure measurement in cmH
2
O.
3.
Set
Pressure Limit
to 20 cmH
2
O, and
t lnsp
to 3.0 seconds.
4.
Verify that the value of Ppeak displayed on the
Numeric
screen, the peak pressure measured by the
calibration analyzer, and the
Pressure Limit
setting are within ±10% of each other. Document the results
on the record sheet.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......