![background image](http://html1.mh-extra.com/html/covidien/newport-e360/newport-e360_service-manual_2673907039.webp)
Overhaul Procedures
Service Manual
2-15
2.8.2
Air and Oxygen Inlet Regulator Rebuild
To remove and install the air and oxygen inlet regulator, follow the procedure in section
“
Regulators, Inlet Blocks, and Flow Sensor Block
.”
Figure 2-12.
Inlet Regulator Rebuild
Note:
Inlet regulators must be outside of the ventilator for disassembly and reassembly.
To install the air and oxygen inlet regulators rebuild kits, refer to
and perform the
following steps:
1.
Unscrew the brass cover from the regulator (no tool should be needed).
2.
Remove the spring, diaphragm, diaphragm washer, seat with o-ring, small spring, and ball.
3.
Using the parts from the regulator rebuild kit (RRK1800M), replace the used parts by reversing the
disassembly.
1
Cover
4
Diaphragm
2
Slip ring
5
O-ring
3
Spring
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......