Component Removal and Replacement
4-18
Service Manual
Figure 4-13.
Pneumatics Panel
6.
Remove the yellow tube and the blue tube from the air and oxygen PSOL valves by disassembling the
luer lock connectors.
7.
While supporting the pneumatics panel, slide the panel away from the ventilator.
8.
To reinstall the panel assembly and components, perform the following steps:
a.
Supporting the pneumatics panel, align the air and oxygen flow sensors with the flow sensor
manifold block.
b.
Continuing to provide support, secure two screws on top for the pneumatics panel (once secured,
it is safe to release support).
c.
Reverse steps 2 through 6.
1
Screws (two of four securing PNL2102M)
Summary of Contents for Newport e360